To achieve a quality neutron radiographic image in a relatively short exposure time, the neutron radiography beam must be of good quality and relatively high neutron flux. Characterization of a neutron radiography beam, such as determination of the image quality and the neutron flux, is vital for producing quality radiographic images and also provides a means to compare the quality of different neutron radiography facilities. This paper provides a characterization of the radial and tangential neutron radiography beamlines at the Tehran research reactor. Thiswork includes determination of the facilities category according to the American Society for Testing and Materials (ASTM) standards, and also uses the gold foils to determine the neutron beam flux. The radial neutron beam is a Category I neutron radiography facility, the highest possible quality level according to the ASTM. The tangential beam is a Category IV neutron radiography facility. Gold foil activation experiments show that the measured neutron flux for radial beamline with length-to-diameter ratio (L/D) = 150 is 6:1 × 10~6 n cm~(-2) s~(-1) and for tangential beamline with (L/D)= 115 is 2:4 × 10~4 n cm~(-2) s~(-1).
机译:为了在相对较短的暴露时间内实现质量的中子X光摄影图像,中子X光片束必须具有良好的质量和相对较高的中子通量。中子X射线照相光束的表征,例如测定图像质量和中子通量,对于产生质量射线照相图像至关重要,并且还提供了一种比较不同中子X光摄影设施质量的方法。本文提供了德黑兰研究反应器上径向和切向中子X射线照相光束的表征。这项工作包括根据美国测试和材料学会(ASTM)标准确定设施类别的确定,还使用金箔来确定中子束通量。径向中子束是I类中子X射线照相设施,根据ASTM的最高质量水平。切向光束是IV类中子射线照相设施。金箔激活实验表明,长度与直径比(l/d)= 150为6:1×10〜6 n cm〜(-2)S〜(-1)的径向光束线的测得的中子通量(-1) (l/d)= 115为2:4×10〜4 n cm〜(-2)s〜(-1)的切向光束线。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号