Room, Suite Scale, Class III Biological Safety Cabinet, and Sensitive Equipment Decontamination and Validation Using Gaseous Chlorine Dioxide

机译:房间,套房秤,III级生物安全柜,以及使用二氧化氯气对敏感设备的净化和验证

获取原文
获取原文并翻译 | 示例
           

摘要

The Tufts New England Regional Biosafety Laboratory facility has been using chlorine dioxide gas for >5 years, and > 100 decontaminations have been done at the room level and on a Class III cabinet. The rooms have ranged from animal holding rooms containing animal racks (housing rodents and rabbits), biosafety cabinets, and computers to biosafety level 3 laboratories containing a variety of equipment (microscopes, biosafety cabinets, centrifuges, incubators, real-time polymerase chain reaction machines, etc). For a biosafety level 3 facility, the equipment is stainless steel where possible, but there is a variety of materials, including electronics and other sensitive equipment. No corrosion has been experienced on any equipment or surface despite repeated decontaminations. Even the most sensitive equipment has not experienced any ill effects. As a test, when decontaminations were started, a low-cost laptop computer was moved from room to room to a Class III cabinet, decontaminating it every time in each chamber. After 35 repeated exposures, it was still functioning with no issues. It is still in use in one of the animal rooms and gets decontaminated along with the room. The Class III cabinet has not shown any traces of corrosion on the stainless-steel interior, nor has the stainless-steel inhalation exposure system kept inside the cabinet experienced any ill effects.
机译:塔夫茨大学新英格兰地区生物安全实验室设施使用二氧化氯气体的时间已经超过5年,并且已经在房间水平和III级机柜上进行了100多次净化。房间的范围从包含动物架子(啮齿动物和兔子)的动物收容室,生物安全柜和计算机到包含各种设备(显微镜,生物安全柜,离心机,孵化器,实时聚合酶链反应机)的生物安全3级实验室。等)。对于3级生物安全设施,在可能的情况下,设备为不锈钢,但材料多种多样,包括电子设备和其他敏感设备。尽管重复进行了去污,但在任何设备或表面上都未发生腐蚀。即使是最敏感的设备也没有遭受任何不良影响。作为测试,当开始进行消毒时,将一台低成本便携式计算机从一个房间移到另一个房间,放置在III级机柜中,每次在每个隔间中对其进行净化。经过35次重复曝光后,它仍然可以正常运行。它仍在其中一间动物室中使用,并随该室一起被净化。 III类机柜在不锈钢内部没有发现任何腐蚀痕迹,柜内的不锈钢吸入暴露系统也没有任何不良影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号