Selective flow injection (FI) procedures were developed for the determination of organic nitrogen compounds, from which ammonia can enzymatically be released. The generated ammonia is detected after gasdialytic separation and derivatization with a totally aqueous o-phthalaldehyde (OPA) reagent. Ammonia can be determined in the range between 10(-7) and 10(-3) M and between 10(-6) and 10(-3) M after the gasdialytic separation. Immobilized amidohydrolases and amino acid oxidases are applied to determine urea, L-glutamine, L-glutamate, L-lysine and L-alanine. An automated four-channel FIA set-up was developed for the sequential determination of ammonia, glutamine, glutamate and alanine. A gas dialytic procedure for the quantitative elimination of ammonia from sample solutions was developed. Up to 15 complete cycles of analysis can be performed per hour. [References: 36]
机译:开发了用于确定有机氮化合物的选择性流动注射(FI)程序,可以从其中酶解释放氨。气体渗析分离并用完全含水的邻苯二甲醛(OPA)试剂衍生化后,检测到生成的氨。氨可以在气体渗析分离后确定在10(-7)和10(-3)M之间以及10(-6)和10(-3)M之间的范围内。应用固定化的酰胺水解酶和氨基酸氧化酶测定尿素,L-谷氨酰胺,L-谷氨酸,L-赖氨酸和L-丙氨酸。开发了一种自动四通道FIA装置,用于顺序测定氨,谷氨酰胺,谷氨酸和丙氨酸。开发了用于从样品溶液中定量去除氨的气体渗析程序。每小时最多可以执行15个完整的分析周期。 [参考:36]