首页> 外文期刊>Analytical chemistry >An ICP-OES Method with 0.2% Expanded Uncertainties for the Characterization of LiAlO_2
【24h】

An ICP-OES Method with 0.2% Expanded Uncertainties for the Characterization of LiAlO_2

机译:不确定度为0.2%的ICP-OES方法用于表征LiAlO_2

获取原文
获取原文并翻译 | 示例
           

摘要

An improved inductively coupled plasma-optical emission specirometry (ICP-OES) method’ has been applied to the determination of U and Al mass fractions and the U/Al amount-of-substance ratio in representative samples of hAlO2. This ICP-OES method has uncertainty on the order of 0.2%,2,3 comparable to the best analytical meth-ods. This method is based on several strategies, which are detailed in this work. The mean measured mass fractions of U and Al in eight samples were 010151 ±0.00016 (+0.16%) and 0A1068 + 0.00056 (+0.14%), and the mean U/Al amount-of-substance ratio was 0.9793 + 0.0017 (±0.17%). The uncertainty is dominated by sample handling and heterogeneity—about a factor of 2 larger than the ICP-OES instrumental uncertainties, which were 0.04% for Al and 0.07% for U.
机译:一种改进的电感耦合等离子体发射光谱法(ICP-OES)已用于测定hAlO2代表性样品中的U和Al质量分数以及U / Al量比。与最佳分析方法相比,这种ICP-OES方法的不确定度约为0.2%,2、3。该方法基于几种策略,本工作将对此进行详细介绍。八个样品中U和Al的平均测量质量分数为010151±0.00016(+ 0.16%)和0A1068 + 0.00056(+ 0.14%),并且U / Al的平均物质含量比为0.9793 + 0.0017(±0.17) %)。不确定性主要由样品处理和异质性决定-大约比ICP-OES仪器不确定性大2倍,ICP-OES的仪器不确定性对Al为0.04%,对于U为0.07%。