A modern electrothermal atomic absorption spectrometer with transversely heated graphite tube and a solid sampling system based on the boat technique was used for the development of a method for the direct determination of Ca, Co, Cr, Cu, Fe, K, Mg, Mn, Na, Ni, and Zn in high-purity tungsten powders. The high background originating from tungsten oxide was eliminated by using hydrogen as purge gas during pyrolysis. Quantification of all elements could be performed by using calibration curves measured with aqueous standard solutions. The analyte contents determined were between 0.02 (Zn) and 34.9 (Fe) μg/g. The results were in good agreement with those obtained in analysis of sample digests, and they were compared also with results provided by other methods in other laboratories. Owing to an almost complete freedom of blank and unusually high sample amounts applied (5-100 mg), extremely low limits of detection, in the range between 0.01 (Mg) and 4 (Fe) ng/g, could be achieved.
机译:使用具有横向加热石墨管和基于舟技术的固体采样系统的现代电热原子吸收光谱仪,开发了一种直接测定Ca,Co,Cr,Cu,Fe,K,Mg,Mn的方法。高纯度钨粉中的Na,Ni和Zn。通过在热解过程中使用氢气作为吹扫气体消除了源自氧化钨的高本底。所有元素的定量都可以通过使用标准水溶液测定的校准曲线来进行。确定的分析物含量在0.02(Zn)和34.9(Fe)μg/ g之间。结果与分析样品消化液中获得的结果非常吻合,并将它们与其他实验室中其他方法提供的结果进行了比较。由于空白几乎完全自由,并且所施加的样品量异常高(5-100 mg),因此可以实现0.01(Mg)到4(Fe)ng / g的极低检测限。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号