首页> 外文期刊>Analytical chemistry >Determination of Selenium(IV) by a Photooxidized 3,3'-Diaminobenzidine/Perfluorinated Polymer Mercury Film Electrode
【24h】

Determination of Selenium(IV) by a Photooxidized 3,3'-Diaminobenzidine/Perfluorinated Polymer Mercury Film Electrode

机译:光氧化的3,3'-二氨基联苯胺/全氟聚合物汞膜电极测定硒(IV)

获取原文
获取原文并翻译 | 示例
           

摘要

A new, and easily fabricated, chemically modified elec-Irode for the determination of selenium(lV) was examined by cathodic square-wave stripping voltammetry. This new electrode consisted of an anion-exchange perfluorinated polymer (Tosfiex) coated thin mercury film electrode containing photooxidized 3,3’-diaminobenzidine (ODAB). The coating solution of Tosfiex and ODAB was spin-coated on a glassy carbon electrode followed by electroplating of a thin film of mercury. During the preconcentration, ODAB was reduced electrochemically and selenium was accumulated simultaneously onto the electrode by inter-acting with the reduced ODAB. After a 5-min preconcen-tration period, linear response was observed from 0.5 to 50 ppb selenium, and the detection limit was 0.1 ppb. The proposed method does not require a darkened room, which was required in many of the previous methods involving 3,3’-diaxninobenzidine. In addition, the resis-tance to interference from surface-active compounds was improved by incorporating Tosfiex in the film.
机译:通过阴极方波溶出伏安法检测了一种易于测定的新型化学修饰的用于测定硒(IV)的电极。这种新电极由阴离子交换全氟聚合物(Tosfiex)涂层的薄汞膜电极组成,该电极含有光氧化的3,3'-二氨基联苯胺(ODAB)。将Tosfiex和ODAB的涂布溶液旋涂在玻璃碳电极上,然后电镀汞薄膜。在预浓缩过程中,ODAB电化学还原,硒与还原的ODAB相互作用,同时将硒累积在电极上。在预浓缩5分钟后,观察到硒从0.5到50 ppb线性响应,检出限为0.1 ppb。所提出的方法不需要变暗的房间,而在以前的许多涉及3,3'-二氮杂联苯胺的方法中都需要这样做。另外,通过在薄膜中掺入Tosfiex可以提高抗表面活性化合物干扰的能力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号