The Validity and Reliability between Automated Oscillometric Measurement of Ankle-Brachial Index and Standard Measurement by Eco-Doppler in Diabetic Patients with or without Diabetic Foot

机译:患有糖尿病患者的腹腔 - 肱骨臂臂指数的自动化示波测量与糖尿病患者中的标准测量的有效性和可靠性

获取原文
获取原文并翻译 | 示例
           

摘要

Objective. To evaluate the concordance between oscillometric ABI and standard Doppler ABI in diabetic Chinese patients with or without diabetic foot. Methods. 230 consecutive diabetic patients (n=459 limbs) were included. The right and left ABIs were determined with both devices by the same investigator. The concordance and agreement were assessed by kappa index and the Bland-Altman method. Results. The average Doppler ABI was 1.003?±?0.286 on the right and 0.990?±?0.287 on the left, while oscillometric ABI was 1.002?±?0.332 and 0.993?±?0.319, which had no significance. The average time for oscillometric ABI was 8.600 versus 16.980 minutes for Doppler ABI (p<0.001). There was good agreement between the two measurements, with a kappa value of 0.869 on the right and 0.919 on the left. Regarding the Doppler ABI as the gold standard, the accuracy, sensitivity, specificity, +LR, and ?LR of oscillometric ABI reached 95.22%, 94.34%, 95.48%, 20.873%, and 0.059% on the right. For the left, it was 96.94%, 96.43%, 97.11%, 33.364%, and 0.036%. Conclusions. The oscillometric measurement is a reliable, convenient, and less time-consuming alternative to standard Doppler ABI in patients. It should be widely used for PAD detection.
机译:客观的。评估糖尿病患者患者的仿制ABI和标准多普勒ABI之间的一致性,患有糖尿病患者的糖尿病患者。方法。包括连续230名连续糖尿病患者(n = 459四肢)。通过同一调查员使用两种设备确定右侧和左臂。 Kappa指数和Bland-Altman方法评估了一致性和协议。结果。平均多普勒ABI为1.003?±0.286左右,左侧0.990?±0.287,而振荡ABI为1.002?±0.332和0.993?±0.319,其没有意义。多普勒ABI的平均散体ABI的平均时间为8.600分钟(P <0.001)。两种测量之间有良好的一致性,右κ值为0.869,左侧0.919。关于多普勒ABI作为黄金标准,精度,敏感性,特异性,+ LR和φ1R均达到95.22%,94.34%,95.48%,20.873%和0.059%。左侧,96.94%,96.43%,97.11%,33.364%和0.036%。结论。示波器测量是患者标准多普勒ABI的可靠,方便,少耗时。它应该广泛用于垫检测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号