摘要

Monte Carlo simulations were used to optimize the geometry of a mammography anti-scatter linear grid to achieve minimum scatter-to-primary ratio (SPR) for different X-ray tube voltages. A single optimum design of the grid with 0.9mm septa height, 12microm septa thickness and 100microm interspace thickness was found for breast phantom thicknesses between 30 and 80mm. The optimal grid has 0.153-0.330 scatter-to-primary ratio, a Bucky factor (BF) less than 2.5 and a contrast improvement factor (CIF) of 1.3.
机译:Monte Carlo模拟用于优化乳房X线摄影抗散射线性电网的几何形状,以实现不同X射线管电压的最小散射对初级比(SPR)。 发现具有0.9mm隔膜高度,12mmrom Secta厚度和100microm间隙厚度的栅格的单一最佳设计,用于30至80mm之间的乳房幻影厚度。 最佳电网具有0.153-0.330散射到初级比率,抗皱因子(BF)小于2.5,对比度改善因子(CIF)为1.3。