IN OUR STUDY WE ASSESSED PREOPERATIVE CATARACT grades in manual and femtosecond laser-assisted cataract surgery and evaluated its influence on effective phacoemulsification times. For phacoemulsiflcation the Infinity System was used (Alcon, Fort Worth, Texas, USA). The effective phacoemulsification time (EPT) is determined by multiplying the phacoemulsification time (seconds) with the average phacoemulsification power (percent).2 In contrast to EPT, the cumulative dissipated energy (CDE) reflects the total U/S energy in footpedal position 3 and includes phaco and torsional times. At present, no phaco platform offers surgeons a direct or standard measure of the energy that is placed into the eye during cataract surgery in units such as Joules. As mentioned by Drs Packer and Solomon, the Infinity system includes the OZil torsional technology, which features both longitudinal and torsional ultrasound delivery. OZil torsional technology produces a side-to-side motion of the phaco tip to break up the nucleus by shearing.
机译:在我们的研究中,我们评估了手动和飞秒激光辅助性白内障手术的术前CATARACT等级,并评估了其对有效超声乳化时间的影响。对于超声乳化术,使用无限系统(美国德克萨斯州沃思堡的爱尔康公司)。有效超声乳化时间(EPT)是通过将超声乳化时间(秒)乘以平均超声乳化力(百分比)来确定的。2与EPT相比,累积消散能量(CDE)反映了足底位置的总U / S能量3并包括晶状体和扭转时间。目前,没有任何phaco平台可为外科医生提供直接或标准的白内障手术中焦耳等单位进入眼内的能量的量度。正如Packer博士和所罗门博士所提到的,Infinity系统包括OZil扭转技术,该技术具有纵向和扭转超声传递功能。 OZil扭转技术产生了phaco尖端的左右运动,以通过剪切使核破裂。