Purpose To determine the incidence of fracture and migration of the Celect inferior vena cava (IVC) filter at a single tertiary-care center. Materials and Methods Images obtained after placement of Celect IVC filters between August 2007 and January 2013 were retrospectively reviewed for evidence of device fracture and migration. The probability of filter fracture at 6 and 12 months after filter placement was estimated, and a 95% upper confidence bound was constructed. Results A total of 741 consecutive Celect filters were placed, and the following studies were subsequently evaluated: 301 abdominal radiographs, 282 abdominal computed tomography (CT) scans, 154 cavagrams, 19 lumbar spine radiographs, 462 chest radiographs, 108 chest CT scans, and 31 pelvic radiographs. The mean imaging follow-up interval was 5.78 months ± 8.80 (range, 0.01-56.03 mo). There was no filter fracture or en-bloc migration outside the IVC. Upper 95% confidence bounds for the probability of fracture were 1.8% at 6 months and 4.3% at 12 months. Intracaval filter migration of more than 2 cm (mean, 2.43 ± 0.12 cm; range, -2.54 to 2.99 cm) was seen in 3.9% of patients (21 of 534; 12 cranial and nine caudal). Conclusions No fracture of a Celect filter was observed in this large cohort over a relatively short follow-up interval. The results also suggest that the Celect filter has an adequate securing mechanism to prevent distant en-bloc or significant intracaval migration.
机译:目的,确定在单个第三级护理中心的晶体下腔静脉(IVC)过滤器的断裂和迁移的发生率。在2007年8月至2013年1月在2013年8月在2007年8月到2013年1月在2013年至2013年于2013年期间获得的图像和方法进行回顾性地审查了装置骨折和移民的证据。估计过滤放置后6和12个月的过滤裂缝概率,构建了95%的上置信度。结果总共741个连续的晶体过滤器,随后进行了以下研究:301腹部射线照相,282腹部计算断层扫描(CT)扫描,154个Cavagrams,19腰脊柱射线照片,462胸射线照片,108个胸部X型扫描,和31骨盆射线照片。平均成像后续间隔为5.78个月±8.80(范围,0.01-56.03 mo)。 IVC外部没有过滤骨折或en-Bloc迁移。骨折概率的高95%置信度偏重为1.8%,在6个月内为12个月,4.3%。在3.9%的患者中,腔内过滤器迁移超过2cm(平均,2.43±0.12厘米;范围,-2.54至2.99厘米)(534分; 12颅颅和九个尾部)。结论在这种大队列中没有在相对短的后续间隔内观察到晶体滤波器的骨折。结果还表明晶体化过滤器具有足够的固定机制,以防止遥远的en-Bloc或显着的腔内迁移。

著录项

  • 来源
  • 作者单位

    Section of Interventional Radiology Imaging Institute Cleveland Clinic 9500 Euclid Ave.;

    Department of Interventional Radiology Suzhou Municipal Hospital Nanjing Medical University;

    Department of Quantitative Health Sciences Cleveland Clinic 9500 Euclid Ave. Cleveland OH 44195;

    Section of Interventional Radiology Imaging Institute Cleveland Clinic 9500 Euclid Ave.;

    Section of Interventional Radiology Imaging Institute Cleveland Clinic 9500 Euclid Ave.;

    Section of Interventional Radiology Imaging Institute Cleveland Clinic 9500 Euclid Ave.;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 放射医学;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号