Abstract Introduction Two methods were compared for evaluating the sigma metrics of clinical biochemistry tests using two different allowable total error (TEa) specifications. Materials and methods The imprecision (CV%) and bias (bias%) of 19 clinical biochemistry analytes were calculated using a trueness verification proficiency testing (TPT)‐based approach and an internal quality control data inter‐laboratory comparison (IQC)‐based approach, respectively. Two sources of total allowable error (TEa), the Clinical Laboratory Improvement Amendments of 1988 (CLIA '88) and the People's Republic of China Health Industry Standard (WS/T 403‐2012), were used to calculate the sigma metrics (σ CLIA, σ WS/T ). Sigma metrics were calculated to provide a single value for assessing the quality of each test based on a single concentration level. Results For both approaches, σ CLIA ??σ WS/T in 18 out of 19 assays. For the TPT‐based approach, 16 assays showed σ CLIA ??3, and 12 assays showed σ WS/T ??3. For the IQC‐based approach, 19 and 16 assays showed σ CLIA ??3 and σ WS/T ??3, respectively. Conclusions Both methods can be used as references for calculating sigma metrics and designing QC schedules in clinical laboratories. Sigma metrics should be evaluated comprehensively by different approaches.
机译:摘要使用两种不同允许的总误差(茶)规格进行比较两种方法,以评估临床生物化学测试的SIGMA度量。材料和方法使用真实的验证熟练程度测试(CTPT)的方法计算了199个临床生物化学分析物的不精确(CV%)和偏差(偏差%)和实验室内部质量控制数据(IQC)的方法, 分别。两个总允许误差(茶)的来源,1988年(Clia '88)和中华人民共和国卫生行业标准(WS / T 403-2012)的临床实验室改善修正案(WS / T 403-2012)用于计算Sigma指标(ΣClia ,σws/ t)。计算Sigma度量标准,以提供单一值,用于基于单浓度水平评估每个测试的质量。两种方法的结果,σClia?σws/ t在19个测定中18个。对于基于TPT的方法,16个测定显示σClia?&β3和12个测定显示σws/t≤α3。对于基于IQC的方法,19和16测定显示σClia?&Δ3和σws/ t?&Δ3。结论这两种方法都可以用作计算SIGMA指标和在临床实验室中设计QC计划的参考。 SIGMA指标应全面评估不同的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号