Objectives: Significantly elevated levels of neurofilament light chain (NfL) and phosphorylated neurofilament heavy chain (pNfH) have been described in the blood and cerebrospinal fluid (CSF) of amyotrophic lateral sclerosis (ALS) patients. The aim of this study was to evaluate the analytical performance of different neurofilament assays in a round robin with 10 centers across Europe/U.S.Methods: Serum, plasma and CSF samples from a group of five ALS and five neurological control patients were distributed across 10 international specialist neurochemical laboratories for analysis by a range of commercial and in-house neurofilament assays. The performance of all assays was evaluated for their ability to differentiate between the groups. The inter-assay coefficient of variation was calculated where appropriate from sample measurements performed across multiple laboratories using the same assay.Results:All assays could differentiate ALS patients from controls in CSF. Inter-assay coefficient of variation of analytical platforms performed across multiple laboratories varied between 6.5% and 41.9%.Conclusions:This study is encouraging for the growing momentum toward integration of neurofilament measurement into the specialized ALS clinic. It demonstrates the importance of 'round robin' studies necessary to ensure the analytical quality required for translation to the routine clinical setting. A standardized neurofilament probe is needed which can be used as international benchmark for analytical performance in ALS.
机译:目的:在肌营养的侧链硬化(ALS)患者的血液和脑脊液(CSF)中描述了神经膜轻链(NFL)和磷酸化神经膜重链(PNFH)的显着升高。本研究的目的是评估欧洲/ USMethods的10个中心的循环罗宾中不同神经丝测定的分析性能:来自一组五种Als和五个神经控制患者的血清,血浆和CSF样品分布在10个国际上一系列商用和内部神经丝测定分析专家神经化学实验室。评估所有测定的性能,以便它们区分组的能力。计算各种间变化系数,其中使用相同的assay横跨多个实验室进行的样品测量来计算。结果:所有测定可以区分ALS患者在CSF中的对照。在多个实验室中进行的分析平台的分析平台的间际变化变化在6.5%和41.9%之间。结论:本研究令人鼓舞的是,在专业的ALS诊所中往往涌入神经丝测量的势头。它展示了“循环罗宾”研究所需的重要性,以确保翻译到常规临床环境所需的分析质量。需要标准化的神经丝探针,可用作ALS中的分析性能的国际基准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号