Accuracy and reliability of fetal heart rate monitoring using maternal abdominal surface electrodes

机译:使用孕妇腹部表面电极监测胎儿心率的准确性和可靠性

获取原文
获取原文并翻译 | 示例
           

摘要

Objective. Compare the accuracy and reliability of fetal heart rate identification from maternal abdominal fetal electrocardiogram signals (ECG) and Doppler ultrasound with a fetal scalp electrode. Design. Prospective open method equivalence study. Setting. Three urban teaching hospitals in the Northeast United States. Sample. 75 women with normal pregnancies in labor at >37 weeks of gestation. Methods. Three fetal heart rate detection methods were used simultaneously in 75 parturients. The fetal scalp electrode was the standard against which abdominal fetal ECG and ultrasound were judged. Main outcome measures. The positive percent agreement with the fetal scalp electrode indicated reliability. Bland-Altman analysis determined accuracy. The confusion rate indicated how frequently the devices tracked the maternal heart rate. Results. Positive percent agreement was 81.7 and 73% for the abdominal fetal ECG and ultrasound, respectively (p = 0.002). The abdominal fetal ECG had a lower root mean square error than ultrasound (5.2 vs. 10.6 bpm, p < 0.001). The confusion rate for ultrasound was 20-fold higher than for abdominal ECG (8.9 vs. 0.4%, respectively, p < 0.001). Conclusion. Compared with the fetal scalp electrode, fetal heart rate detection using abdominal fetal ECG was more reliable and accurate than ultrasound, and abdominal fetal ECG was less likely than ultrasound to display the maternal heart rate in place of the fetal heart rate.
机译:目的。通过胎儿头皮电极比较根据孕妇腹部胎儿心电图信号(ECG)和多普勒超声检查胎儿心率的准确性和可靠性。设计。前瞻性开放方法等效性研究。设置。美国东北部的三所城市教学医院。样品。 75名在妊娠> 37周时分娩正常的妇女。方法。在75例产妇中同时使用了三种胎儿心率检测方法。胎儿头皮电极是判断腹部胎儿心电图和超声的标准。主要观察指标。与胎儿头皮电极的正百分比一致性表明可靠性。 Bland-Altman分析确定了准确性。混乱率表明设备跟踪孕产妇心率的频率。结果。腹部胎儿心电图和超声检查的阳性一致性百分比分别为81.7和73%(p = 0.002)。腹部胎儿心电图的均方根误差低于超声检查(5.2 vs. 10.6 bpm,p <0.001)。超声的混乱率比腹部ECG高20倍(分别为8.9%和0.4%,p <0.001)。结论。与胎儿头皮电极相比,使用腹部胎儿心电图检测胎儿的心率比超声更可靠,更准确,而腹部胎儿心电图比超声显示孕妇心率代替胎儿心率的可能性更低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号