BACKGROUND: Controversy still exists as to whether to mount the tibial bearing on a metal baseplate or not. Furthermore, the tibial component can be either horizontally or completely cemented. We evaluated metal backing versus all-polyethylene using horizontal cementing technique. PATIENTS AND METHODS: We randomized 40 patients with knee arthrosis (40 knees) to 2 groups: cemented total knee arthroplasty (AGC, Biomet) using either all-polyethylene (AP) or metal-backed (MB) tibial components (n = 20 for each group). All tibial implants had a total thickness of 8 mm. We used proximal cement-ing, including only the horizontal part of the tibia with avoidance of the stem-bone interface. RESULTS: The positioning of the implants, as judged by the HKA angle, showed an average neutral alignment in both groups. Preoperatively and after 2 years, there was no statistically significant difference between the HSS scores in both groups (p = 0.6 and 0.4). After 2 years, the metal-backed components had rotated more around the longitudinal axis, median 0.5 degrees vs. 0.2 degrees (internal or external rotation, p = 0.002), and showed increased maximum total point motion, median 1.0 vs. 0.4 mm (maximum migration or MTPM, p = 0.003). INTERPRETATION: Our study indicates that better fixation can be obtained with the all-polyethylene design if proximally cemented thin components are used.
机译:背景:关于是否将胫骨轴承安装在金属底板上仍存在争议。此外,胫骨组件可以水平或完全固定。我们使用水平胶凝技术评估了金属背衬与全聚乙烯的对比。患者与方法:我们将40例膝关节关节炎(40膝)患者随机分为两组:使用全聚乙烯(AP)或金属支持(MB)胫骨组件进行全骨水泥置换(AGC,Biomet)(n = 20每组)。所有胫骨植入物的总厚度为8毫米。我们使用了近端骨水泥固定术,仅包括胫骨的水平部分,避免了骨干界面。结果:根据HKA角度判断,植入物的位置在两组中均显示中性对齐。术前和术后2年,两组的HSS评分之间无统计学差异(p = 0.6和0.4)。 2年后,金属支撑的组件绕纵轴旋转的更多,中点0.5度比0.2度(内部或外部旋转,p = 0.002),并且最大总点运动增加,中点1.0相对0.4毫米(最大迁移量或MTPM,p = 0.003)。解释:我们的研究表明,如果使用近端粘结的薄部件,则采用全聚乙烯设计可以获得更好的固定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号