首页> 外文期刊>Spectrochimica acta, Part A. Molecular and biomolecular spectroscopy >Simultaneous multi-gas detection between 3 and 4 mu m based on a 2.5-m multipass cell and a tunable Fabry-Perot filter detector
【24h】

Simultaneous multi-gas detection between 3 and 4 mu m based on a 2.5-m multipass cell and a tunable Fabry-Perot filter detector

机译:基于2.5M多峰电池和可调谐法布里 - 珀罗滤波器检测器的3至4μm的同时多气体检测

获取原文
获取原文并翻译 | 示例
           

摘要

We demonstrated a versatile and innovative gas sensing system based on a Fabry-Perot (FP) filter detector, which operates in the spectral range from 3.1 to 4.4 mu m (3226-2273 cm(-1)) with a spectral resolution of 20 nm. The developed sensor system can be used to record the entire spectrum by means of a one-time scan or, alternatively, to access selected spectral regions by using the tunable FP filter detector. A multipass cell with an effective path length of 2.5 m was implemented to improve the detection sensitivity. The spectra of methane, formaldehyde and carbon dioxide were simultaneously measured, with detection limits of 200 ppm, 900 ppm and 20 ppm, respectively. A seven-day continuous measurement for indoor carbon dioxide gas was carried out demonstrating the stability and robustness of the reported sensor system. (C) 2019 Elsevier B.V. All rights reserved.
机译:我们证明了一种基于法布里 - 珀罗(FP)滤波器检测器的通用和创新的气体传感系统,其在光谱范围内从3.1到4.4 mu m(3226-2273cm(-1)),光谱分辨率为20 nm 。 开发的传感器系统可用于通过一次性扫描或替代地,通过使用可调谐FP滤波器检测器来获取所选择的光谱区域来记录整个光谱。 实施有有效路径长度为2.5米的多峰值电池以提高检测灵敏度。 同时测量甲烷,甲醛和二氧化碳的光谱,分别具有200ppm,900ppm和20ppm的检测限。 进行七天的室内二氧化碳气体的连续测量,表明报告的传感器系统的稳定性和稳健性。 (c)2019 Elsevier B.v.保留所有权利。

著录项

  • 来源
  • 作者单位

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Shanxi Univ Inst Laser Spect State Key Lab Quantum Opt &

    Quantum Opt Devices Taiyuan 030006 Shanxi Peoples R China;

    Rice Univ Dept Elect &

    Comp Engn Houston TX 77005 USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 光化学分析法(光谱分析法);
  • 关键词

    Optical gas sensor; Fabry-Perot filter detector; Multi-gas detection;

    机译:光学气体传感器;法布里 - 珀罗过滤器检测器;多气体检测;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号