Falsely elevated cyclosporin and tacrolimus concentrations over prolonged periods of time due to reversible adsorption to central venous catheters

机译:由于对中心静脉导管的可逆吸附,长时间延长了环孢菌素和他克莫司的浓度

获取原文
获取原文并翻译 | 示例
           

摘要

Falsely elevated concentrations of immunosuppressants can be caused by reversible adsorption to central venous catheter (CVC) systems. If undetected, this may lead to dose reduction resulting in underdosage which may even entail graft-versus-host disease or organ rejection. We analyzed the adsorption and release for cydosporine A (CsA) and tacrolimus (Tac) in vitro and in vivo. Four types of CVCs were examined in vitro: two made from poly-urethane (PU), one from silicone and one from PU with an incorporated silver ion-based antimicrobial agent All 26 CVCs analyzed in vitro showed significant reversible adsorption of CsA (n = 13; p = 0.001) and Tac (n = 13; p = 0.001, Wilcoxon signed rank test). Immediately after infusing the drugs, the mean concentrations of 6420 ng/mL of CsA and 250 ng/mL of Tac were measured. Flushing with NaCl lowered the drug release. Besides, blood samples of fifteen patients were taken simultaneously from all lumina of the CVC and via venipuncture. The samples from contaminated lumina showed the mean elevations by a factor of 11 for CsA (n = 12) and 89 for Tac (n = 3). Blood sampling for immunosuppressant monitoring should thus never be performed from lumina previously used for infusing the drug even after prolonged periods of time and extensive rinsing.
机译:免疫抑制剂浓度错误升高可能是由于对中心静脉导管(CVC)系统的可逆吸附引起的。如果未被发现,这可能导致剂量减少,导致剂量不足,甚至可能导致移植物抗宿主疾病或器官排斥。我们在体外和体内分析了环孢霉素A(CsA)和他克莫司(Tac)的吸附和释放。在体外检查了四种类型的CVC:两种由聚氨酯(PU)制成,一种由硅树脂制成,一种由具有银离子基抗菌剂的PU制成,在体外分析的所有26种CVC均显示出可逆的CsA吸附性(n = 13; p = 0.001)和Tac(n = 13; p = 0.001,Wilcoxon符号秩检验)。注入药物后,立即测量平均浓度为6420 ng / mL的CsA和250 ng / mL的Tac。用氯化钠冲洗可降低药物释放。此外,从CVC的所有腔内并通过静脉穿刺同时采集了15名患者的血样。来自受污染的氧化铝的样品显示CsA的平均升高幅度为11(n = 12),Tac的平均升高幅度为89(n = 3)。因此,即使经过长时间和广泛的冲洗,也绝不能从以前用于注入药物的腔中进行血液采样以进行免疫抑制剂监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号