首页> 外文期刊>Journal of the mechanical behavior of biomedical materials >Failure analysis of the fractured wires in sternal perichronal loops.
【24h】

Failure analysis of the fractured wires in sternal perichronal loops.

机译:胸骨颞骨环中骨折线的失效分析。

获取原文
获取原文并翻译 | 示例
           

摘要

We report failure analysis of sternal wires in two cases in which a perichronal fixation technique was used to close the sternotomy. Various characteristics of the retrieved wires were compared to those of unused wires of the same grade and same manufacturer and with surgical wire specifications. In both cases, wire fracture was un-branched and transgranular and proceeded by a high cycle fatigue process, apparently in the absence of corrosion. However, stress anlysis indicates that the effective stress produced during strong coughing is lower than the yield strength. Our findings suggest that in order to reduce the risk for sternal dehiscence, the diameter of the wire used should be increased.
机译:我们报告了胸骨切开术的两种情况下胸骨钢丝失败分析的两种情况。将回收的线材的各种特性与同等级,相同制造商且具有手术线规格的未使用线材进行比较。在这两种情况下,金属丝的断裂都是无分支的和横穿的,并且通过高循环疲劳过程进行,显然没有腐蚀。但是,应力分析表明,强咳嗽期间产生的有效应力低于屈服强度。我们的发现表明,为了减少胸骨裂开的风险,应增加所用导线的直径。