Organ doses can be estimated from the computed tomography (CT) dose index for cone-beam CT on radiotherapy equipment

机译:可以从放射治疗设备上的锥形束CT的计算机断层扫描(CT)剂量指数估算器官剂量

获取原文
获取原文并翻译 | 示例
           

摘要

Cone beam computed tomography (CBCT) systems are fitted to radiotherapy linear accelerators and used for patient positioning prior to treatment by image guided radiotherapy (IGRT). Radiotherapists' and radiographers' knowledge of doses to organs from CBCT imaging is limited. The weighted CT dose index for a reference beam of width 20 mm (CTDIw,ref) is displayed on Varian CBCT imaging equipment known as an On-Board Imager (OBI) linked to the Truebeam linear accelerator. This has the potential to provide an indication of organ doses. This knowledge would be helpful for guidance of radiotherapy clinicians preparing treatments. Monte Carlo simulations of imaging protocols for head, thorax and pelvic scans have been performed using EGSnrc/BEAMnrc, EGSnrc/DOSXYZnrc, and ICRP reference computational male and female phantoms to derive the mean absorbed doses to organs and tissues, which have been compared with values for the CTDIw,ref displayed on the CBCT scanner console. Substantial variations in dose were observed between male and female phantoms. Nevertheless, the CTDIw,ref gave doses within +/- 21% for the stomach and liver in thorax scans and 2 x CTDIw,ref can be used as a measure of doses to breast, lung and oesophagus. The CTDIw,ref could provide indications of doses to the brain for head scans, and the colon for pelvic scans. It is proposed that knowledge of the link between CTDIw for CBCT should be promoted and included in the training of radiotherapy staff.
机译:锥束计算机断层扫描(CBCT)系统已安装在放射治疗线性加速器上,并在通过影像引导放射治疗(IGRT)治疗之前用于患者定位。放射治疗师和放射照相师对CBCT成像对器官剂量的了解有限。宽度为20 mm的参考光束的加权CT剂量指数(CTDIw,ref)显示在链接到Truebeam线性加速器的称为板载成像仪(OBI)的Varian CBCT成像设备上。这有可能提供器官剂量的指示。这些知识将有助于指导放射治疗临床医生准备治疗。已使用EGSnrc / BEAMnrc,EGSnrc / DOSXYZnrc和ICRP参考计算的男性和女性人体模型对头部,胸部和骨盆扫描的成像协议进行了蒙特卡洛模拟,以得出器官和组织的平均吸收剂量,并将其与值进行了比较。 CDI CT扫描仪控制台上显示的CTDIw,ref。在男性和女性幻影之间观察到剂量的显着变化。然而,在胸部扫描中,CTDIw,ref对胃和肝的剂量在+/- 21%以内,而2 x CTDIw,ref可以用作对乳房,肺和食道的剂量的量度。 CTDIw,ref可以为头扫描提供给大脑的剂量指示,为盆腔扫描提供给结肠的剂量指示。建议在CTDIw与CBCT之间的联系方面应加深认识,并将其纳入放射治疗人员的培训中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号