首页> 外文期刊>Biomedical sciences instrumentation >Electric field penetration depth of myocardial surface catheters and the measurement of myocardial resistivity.
【24h】

Electric field penetration depth of myocardial surface catheters and the measurement of myocardial resistivity.

机译:心肌表面导管的电场穿透深度和心肌电阻率的测量。

获取原文
获取原文并翻译 | 示例
           

摘要

The in vivo measurement of cardiac conductance in mice offers a method to generate an instantaneous left ventricular volume signal. In order to translate the measured conductance to volume, it is necessary to determine the resistivity of the myocardial tissue. This is done using tetrapolar surface conductance catheters, placed on the surface of the left ventricle. It is important to determine the depth of penetration of the electric field of this surface catheter to ensure that the electric field is confined within the myocardium and does not extend into the left ventricle blood volume. The depth of penetration is experimentally determined by a technique described by Foster et al. [2].
机译:小鼠体内电导的体内测量提供了一种产生瞬时左心室容积信号的方法。为了将测得的电导转换为体积,必须确定心肌组织的电阻率。这是通过使用放置在左心室表面的四极表面电导导管来完成的。确定该表面导管的电场穿透深度很重要,以确保电场被限制在心肌内并且不会扩展到左心室血容量中。渗透深度由Foster等人描述的技术通过实验确定。 [2]。