The aim of this study was to evaluate the volume-of-interest (VOI) technique in the measurement of volume radioactivity and in the differentiation of necrotic sites from residual tumor activity in a phantom. METHODS: PET/CT was performed on a phantom filled with (18)F-FDG solution at different concentrations. The VOI was quantified in 2 sessions to evaluate the VOI measurements as a function of activity concentration in the phantom. Software was used to build the VOI, determine the volume radioactivity of the cylindric inserts (cm(3)), and compare them with their real volumes. The VOI technique was also used to discern the mixed distribution of regions of (18)F-FDG activity from cold regions that represent areas of necrosis without tumor activity. RESULTS: Volumes measured with the VOI technique were similar to the actual volumes of cylinders in the phantom (no statistical differences; P > 0.05 after t test analysis). The diameter of cold inserts correlated positively with the percentage of visualization (P < 0.01); in both sessions, it was possible to visualize 100% of the 12.7-, 11.1-, and 9.5-mm cold rods. CONCLUSION: VOI technique has shown great potential for evaluating volume radioactivity and differentiating hot and cold regions in a phantom; clinical studies should be performed with this technique to evaluate its utility.
机译:这项研究的目的是评估感兴趣的体积(VOI)技术,以测量体积的放射性以及从模型中残留的肿瘤活动中区分坏死部位。方法:PET / CT在以不同浓度的(18)F-FDG溶液填充的体模上进行。在2个阶段中对VOI进行量化,以评估VOI测量值与体模中活动浓度的关系。使用软件来构建VOI,确定圆柱状插入物的体积放射性(cm(3)),并将其与其实际体积进行比较。 VOI技术也被用来区分(18)F-FDG活动区域与寒冷地区的混合分布,寒冷地区代表没有肿瘤活动的坏死区域。结果:使用VOI技术测量的体积与体模中的圆柱体实际体积相似(无统计学差异; t检验后P> 0.05)。冷镶件的直径与可视化百分比呈正相关(P <0.01);在这两个阶段中,都可以可视化100%的12.7、11.1和9.5毫米冷棒。结论:VOI技术已显示出巨大的潜力,可以评估体模中的放射性和区分冷热区域。应使用该技术进行临床研究以评估其效用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号