A simple, lightweight CPAP-delivery device, composed of a three-way stopcock, for the nondependent lung.

机译:一种简单,轻便的CPAP输送装置,由三通旋塞阀组成,用于非依赖性肺部。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: We aimed to introduce a simple, lightweight continuous positive airway pressure (CPAP)-delivery device for the nondependent lung during one-lung ventilation, to investigate how the type of three-way stopcocks, and the compliance and resistance of a test lung affect the relationship between the oxygen flow rate and CPAP level produced, and to examine how the device works in a clinical setting. METHODS: In the test lung study, the bronchial blocker of a Univent tube was connected to a test lung. The effects of oxygen-flow rate, types of three-way stopcocks, and compliance and resistance of the test lung on the CPAP levels were studied. In the clinical study, the lightweight device was used to treat hypoxia in seven patients during one-lung ventilation with the bronchial blocker. RESULTS: In the test lung study, the CPAP level produced by the device was proportional to the oxygen-flow rate, dependent on the type of three-way stopcock used, and independent of the compliance or resistance of the test lung. There was no discrepancy between the plateau pressures of the test lung and the monitoring port of an additional stopcock at any degree of compliance or resistance of the test lung at any oxygen-flow rate. Therefore, the relationship between the oxygen-flow rate and CPAP level can be ensured in advance before application to the lung, with an additional three-way stopcock of which the distal end is occluded. In the clinical study, peripheral oxygen sataration [Formula: see text] improved while the CPAP level ranged from 2.8 to 5.4 cmH(2)O. CONCLUSION: The lightweight CPAP delivery-device can provide variable CPAP levels by adjusting the oxygen-flow rate without real-time monitoring of the pressure.
机译:目的:我们的目的是为单肺通气中的非依赖性肺引入一种简单,轻便的连续气道正压通气装置,以研究三通旋塞的类型以及测试肺的顺应性和阻力影响氧气流速和产生的CPAP水平之间的关系,并检查设备在临床环境中的工作方式。方法:在试验肺研究中,将Univent管的支气管阻滞剂连接到试验肺。研究了氧气流速,三通旋塞阀的类型以及测试肺的顺应性和阻力对CPAP水平的影响。在临床研究中,这种轻巧的设备被用于治疗7例使用支气管阻滞剂进行单肺通气的低氧患者。结果:在试验肺的研究中,该装置产生的CPAP水平与氧气流速成正比,取决于所用的三通旋塞阀的类型,并且与试验肺的顺应性或阻力无关。在任何氧气流速下,在测试肺的顺应性或阻力水平下,测试肺的高原压力与附加旋塞阀的监测口之间均无差异。因此,在施加到肺部之前,可以预先确保氧气流速和CPAP水平之间的关系,并且另外的三通旋塞阀的远端被阻塞。在临床研究中,当CPAP水平在2.8至5.4 cmH(2)O范围内时,外周血氧饱和度得到改善。结论:轻量型CPAP输送装置可通过调节氧气流量来提供可变的CPAP水平,而无需实时监测压力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号