首页> 外文期刊>Journal of Analytical Toxicology >Resolution of an apparent hook effect in roche partner DRI oxycodone immunoassay
【24h】

Resolution of an apparent hook effect in roche partner DRI oxycodone immunoassay

机译:罗氏伴侣DRI羟考酮免疫分析中表观钩效应的解析

获取原文
获取原文并翻译 | 示例
           

摘要

A presumed hook effect in the semiquantitative DRI Oxycodone immunoassay, OXY3S (Cobas Integra, Roche Diagnostics), was investigated in 14 urine samples with gas chromatography/mass spectrometry (GC-MS) >10,000 ng/mL but OXY3S <1,000 ng/mL. These samples included the index case, a false-negative OXY3S result with >75,000 ng/mL oxycodone 1 oxymorphone by GC-MS confirmation. Patient samples needed 2-to 16-fold dilution to obtain the correct OXY3S response. The OXY3S test did not hook at highspiked concentrations of oxycodone, oxymorphone or oxymorphone- 3b-D-glucuronide in drug-free urine. The OXY3S test parameters were replicated in a development channel on the Cobas using DRI Reagents (Microgenics, CA, USA) and were subsequently modified. Delayed sample addition or doubling of Reagent 1 (R1: antibody/substrate/co-factor) yielded maximal immunoassay response (>10,000 ng/mL) in 12 of 14 and 14 of 14 undiluted patient samples, respectively. Supplementation of R1 with substrate alone did not correctly recover oxycodone from any of the samples, while co-factor supplementation resulted a maximal OXY3S response in 13 of 14 samples. The remaining (index) sample could only be corrected by supplemental R1. The semiquantitative utility of the DRI Oxycodone assay is questionable. Although the precise cause of the under-recovery could not be determined, the modification presented permits reliable oxycodone determination at the high concentrations frequently seen in clinical urine samples.
机译:使用气相色谱/质谱(GC-MS)> 10,000 ng / mL,但OXY3S ​​<1,000 ng / mL,对14种尿液样品进行了半定量DRI羟考酮免疫测定OXY3S(Cobas Integra,Roche Diagnostics)的推测钩效应。这些样品包括索引情况,通过GC-MS确认,假阴性OXY3S结果,> 75,000 ng / mL羟考酮1羟吗啡酮。患者样品需要2到16倍稀释才能获得正确的OXY3S反应。在无药尿液中,高浓度的羟考酮,羟吗啡酮或羟吗啡酮-3b-D-葡糖醛酸苷含量较高时,OXY3S测试未引起关注。使用DRI试剂(Microgenics,CA,美国)将OXY3S测试参数复制到Cobas的开发通道中,然后进行修改。延迟添加或加倍试剂1(R1:抗体/底物/辅因子)的样品分别在14个未稀释的患者样品中的14个和14个中的14个中产生了最大的免疫测定响应(> 10,000 ng / mL)。仅使用底物补充R1不能从任何样品中正确回收羟考酮,而辅因子补充在14个样品中的13个中导致最大的OXY3S反应。剩余的(指数)样本只能通过补充R1进行校正。 DRI羟考酮测定法的半定量实用性值得怀疑。尽管无法确定恢复不足的确切原因,但所提出的修改方法可以在临床尿液样本中常见的高浓度下可靠地测定羟考酮。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号