首页> 外文期刊>Dermatologic surgery >Determination of Optimal Parameters for Laser for Nonablative Remodeling with a 1.54 &mgr;m Er:Glass Laser: A Dose-Response Study.
【24h】

Determination of Optimal Parameters for Laser for Nonablative Remodeling with a 1.54 &mgr;m Er:Glass Laser: A Dose-Response Study.

机译:用1.54μmEr:玻璃激光器确定非烧蚀重塑激光器的最佳参数:剂量响应研究。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Nonablative subsurface heating is a challenge for medical laser companies. Among the different lasers proposed today, the erbium-glass (Er:glass) laser combined with contact cooling could be an alternative for nonablative remodeling. OBJECTIVE: To investigate the risk of side effects according to the quality of cooling and number of pulses. METHODS: A clinical evaluation was performed on 10 patients using an Er:glass laser (1.54 &mgr;m) with contact sapphire cooling handpiece (+5 degrees C). Periorbital and perioral areas were treated with a number of pulses increasing gradually from three to eight for a total energy of 24-64 J/cm2. The presence or absence of swelling, crust, blister, and erythema were graded at 7 days. At 3 months postoperatively, hypopigmentation, hyperpigmentation, textural changes, and scars were evaluated clinically. RESULTS: Periorbital and perioral areas respond differently to the number of pulses defining a "safe" clinical threshold. The periorbital site is very sensitive to dermal heating and efficacy of contact cooling; the anatomic features of this body location may explain these differences. CONCLUSION: Selective dermal heating can be achieved with an Er:glass laser coupled with a contact cooling handpiece. The quality of the contact cooling and the number of pulses appear to be important parameters for safety and reproducible clinical results.
机译:背景:非烧蚀性地下加热是医疗激光公司面临的挑战。在当今提出的不同激光器中,结合接触冷却的玻璃(Er:glass)激光器可以作为非烧蚀重塑的替代方法。目的:根据冷却质量和脉冲数调查副作用的风险。方法:使用Er:glass激光(1.54μm)和接触式蓝宝石冷却手机(+5摄氏度)对10例患者进行了临床评估。沿周缘和口周区域处理的脉冲数从3个逐渐增加到8个,总能量为24-64 J / cm2。在第7天对肿胀,结rust,水泡和红斑的存在与否进行分级。术后3个月,对临床上色素沉着,色素沉着,纹理变化和疤痕进行了评估。结果:牙周和口周区域对定义“安全”临床阈值的脉冲数量的反应不同。眼眶周部位对真皮加热和接触冷却功效非常敏感。这种身体位置的解剖特征可以解释这些差异。结论:Er:玻璃激光器与接触式冷却手机可实现选择性的皮肤加热。接触冷却的质量和脉冲数似乎是安全性和可再现的临床结果的重要参数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号