A procedure for the preconcentration and determination of lead in vegetable and biological samples was developed in the continuous mode coupled to a flame atomic absorption spectrometer. Lead is quantitatively preconcentrated in acetic buffer as its diethyldithiocarbamate chelate onto a C-18 minicolumn, placed in the loop of a proportional injector, eluted by a stream of methyl isobutyl ketone and introduced directly into the nebuliser. A detection limit of 3 μg 1~(-1) is obtained using a time-based technique for 2 min preconcentration and an RSD of 3.8% was readily achieved for three measurements of 25 μg Pb 1~(-1). The sample throughput is 24 h~(-1). Using preconcentration times of 10 min an enrichment factor of 189 can be obtained. The continuous flow system was used for some reference sample analysis and the obtained results reveal that the methodology can be easily applied for vegetable and biological sample analysis.
机译:以与火焰原子吸收光谱仪耦合的连续模式开发了蔬菜和生物样品中铅的预浓缩和测定方法。铅在乙酸缓冲液中定量预浓缩,因为其二乙基二硫代氨基甲酸酯与C-18微型色谱柱螯合,置于比例注射器的回路中,由甲基异丁基酮流洗脱并直接引入雾化器中。使用基于时间的技术进行2分钟预浓缩可获得3μg1〜(-1)的检出限,而对于25μgPb 1〜(-1)的三份测量,RSD则很容易达到3.8%。样品通量为24 h〜(-1)。使用10分钟的预浓缩时间,可以获得189的富集系数。连续流系统用于一些参考样品分析,所得结果表明该方法可轻松用于蔬菜和生物样品分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号