Drainage failure because of spontaneous fracture of the peritoneal dialysis catheter.

机译:由于腹膜透析导管自发性骨折而导致排液失败。

获取原文
获取原文并翻译 | 示例
           

摘要

Outflow failure is attributed mostly to mechanical problems such as catheter migration, omental wrapping, constipation, and intraluminal obstruction with fibrin or clot; and it usually occurs shortly after initiation of peritoneal dialysis (PD). Fracture or rupture of the PD catheter is, however, a very unusual cause of drainage failure. We recently experienced a case of early drainage failure because of silicone PD catheter fracture, without previous mechanical or chemical injury, that occurred 1 month after initiation of PD.
机译:流出失败主要归因于机械问题,例如导管迁移,网膜包裹,便秘和纤维蛋白或血凝块引起的腔内阻塞。通常在腹膜透析(PD)开始后不久发生。然而,PD导管的破裂或破裂是引流失败的非常不寻常的原因。我们最近经历了一个案例,该案例是由于硅胶PD导管断裂而导致的早期引流失败,而没有发生机械或化学损伤,发生在PD发作后1个月。