首页> 外文期刊>Soil Science Society of America Journal >A CARBON DIOXIDE FLUX GENERATOR FOR TESTING INFRARED GAS ANALYZER-BASED SOIL RESPIRATION SYSTEMS
【24h】

A CARBON DIOXIDE FLUX GENERATOR FOR TESTING INFRARED GAS ANALYZER-BASED SOIL RESPIRATION SYSTEMS

机译:用于测试基于红外气体分析仪的土壤呼吸系统的二氧化碳通量发生器

获取原文
获取原文并翻译 | 示例
           

摘要

An artificial flux generation device was developed to test the accuracy of a closed-dynamic soil respiration system (LICOR 6400). The device consisted of an enclosed reservoir with a porous top; the reservoir contained a volume of CO2 enriched air, which was monitored by an infrared gas analyzer (IRGA). When the internal CO2 concentration within the reservoir was elevated, diffusion rates through the porous medium were measured by recording changes in CO2 concentration within the reservoir. This diffusion-based artificial flux mimics natural soil respiration, and allows an independent verification of the accuracy of soil respiration measurement systems. We tested a LI-COR 6400 portable photosynthesis system fitted with a 6400-09 soil CO2 flux chamber. On average, this system overestimated high flux rates by 2 to 4% and underestimated low flux rates by 4 to 20% over five independent trials. Soil respiration appeared to be sensitive to boundary layer mixing, and ambient CO2 concentrations.
机译:开发了一种人工通量产生装置,以测试封闭动态土壤呼吸系统(LICOR 6400)的 准确性。该设备由一个封闭的容器构成,该容器具有一个 多孔顶部;储层中含有一定体积的富含CO 2 空气,并通过红外气体分析仪(IRGA)对其进行监控。 当内部CO 升高储层内> 2 的浓度 ,通过记录内部CO 2 浓度的变化来测量通过多孔介质的扩散速率 基于扩散的人工通量模拟自然土壤呼吸, ,并允许独立验证土壤 呼吸测量系统的准确性。我们测试了装有6400-09土壤CO 2 通量室的LI-COR 6400便携式 光合作用系统。 平均而言,该系统高估了在五个独立的 试验中,通量率降低了2至 4%,低通量率降低了4至20%。土壤呼吸似乎对边界 层混合以及环境CO 2 浓度敏感。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号