首页> 外文期刊>Review of Scientific Instruments >Determining time resolution of microchannel plate detectors for electron time-of-flight spectrometers
【24h】

Determining time resolution of microchannel plate detectors for electron time-of-flight spectrometers

机译:确定电子飞行时间谱仪的微通道板探测器的时间分辨率

获取原文
获取原文并翻译 | 示例
           

摘要

The temporal resolution of a 40 mm diameter chevron microchannel plate (MCP) detector followed by a constant fraction discriminator and a time-to-digital converter was determined by using the third order harmonic of 25 fs Ti:sapphire laser pulses. The resolution was found to deteriorate from 200 to 300 ps as the total voltage applied on the two MCPs increased from 1600 to 2000 V. This was likely due to a partial saturation of the MCP and/or the constant fraction discriminator working with signals beyond its optimum range of pulse width and shape.
机译:通过使用25 fs Ti:蓝宝石激光脉冲的三次谐波来确定直径为40毫米的V形V形微通道板(MCP)检测器,然后是恒定分数鉴别器和时间数字转换器的时间分辨率。发现施加在两个MCP上的总电压从1600 V增加到2000 V时,分辨率从200 ps降低到300 ps。这很可能是由于MCP的部分饱和和/或恒定分数鉴别器对超出其范围的信号起作用脉冲宽度和形状的最佳范围。