首页> 外文期刊>Radiological Physics and Technology >An EGSnrc investigation of the air-kerma strength, dose rate constant, and radial dose function of 125I brachytherapy sources
【24h】

An EGSnrc investigation of the air-kerma strength, dose rate constant, and radial dose function of 125I brachytherapy sources

机译:EGSnrc对125 I近距离放射治疗源的气动力强度,剂量率常数和径向剂量函数的研究

获取原文
获取原文并翻译 | 示例
           

摘要

Titanium-encapsulated 125I brachytherapy sources are in use for treatment of the eye, brain, and head and neck region, and for early stage prostate cancer. The photoelectric interaction of 125I photons with titanium encapsulation generates Ti K X-rays (~5 keV). According to the National Institute of Standards and Technology (NIST) 1999 air-kerma strength, S k, standard, these X-rays should be excluded from S k. We used the EGSnrc Monte Carlo code system to calculate the S k (including the contribution of ~5-keV X-rays), dose rate constant, and radial dose function for five different 125I source models. Depending upon the source model, the contribution of 5-keV Ti X-rays to S k varies between 17.1 and 18.7%. Including these X-rays as part of S k would result in underestimation of the dose rate constant by up to 19%. The radial dose functions of the investigated sources are comparable to published studies that are based on an updated photon cross-section dataset.
机译:钛包裹的​​125 I近距离放射治疗源已用于治疗眼,脑和头颈部区域以及早期前列腺癌。 125 I光子与钛封装的光电相互作用产生Ti K X射线(〜5 keV)。根据美国国家标准技术研究院(NIST)1999年的空气比释动能强度S k 标准,应将这些X射线排除在S k 外。我们使用EGSnrc蒙特卡罗代码系统计算了五个不同的125 I源模型的S k (包括〜5-keV X射线的贡献),剂量率常数和径向剂量函数。根据源模型的不同,5-keV Ti X射线对S k的贡献在17.1%和18.7%之间变化。将这些X射线作为S k 的一部分包括在内,会导致剂量率常数低估多达19%。研究来源的径向剂量函数与基于更新的光子截面数据集的已发表研究相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号