首页> 外文期刊>Radiation Protection Dosimetry >MEASUREMENT OF THE MODULATION TRANSFER FUNCTION OF DIGITAL X-RAY DETECTORS WITH AN OPAQUE EDGE-TEST DEVICE
【24h】

MEASUREMENT OF THE MODULATION TRANSFER FUNCTION OF DIGITAL X-RAY DETECTORS WITH AN OPAQUE EDGE-TEST DEVICE

机译:带有不透明边缘测试装置的数字X射线探测器的调制传递函数的测量

获取原文
获取原文并翻译 | 示例
           

摘要

A variant of the edge method for the determination of the pre-sampled modulation transfer function (MTF) of digital X-ray imaging devices has been developed and accepted as the standard method in the novel DQE measurement standard IEC 62220-1. An opaque tungsten edge-test device accomplishes the ideal step-like profile of the incident X rays. The edge spread function is measured over a large region across the edge transition that enables an accurate MTF measurement including the 'low-frequency drop'. The method has been applied to different state-of-the-art X-ray imaging detectors, a computed radiography, a CsI-based indirect and an Se-based direct flat-panel detector. The MTF measurement results will be presented. In contrast to the opaque edge device, the commonly used semi-transparent edge-test devices produce scatter radiation that deteriorates the incident X-ray profile, which leads to a systematic overestimation of the MTF.
机译:已经开发出用于确定数字X射线成像设备的预采样调制传递函数(MTF)的边缘方法的一种变体,并已被接受为新颖的DQE测量标准IEC 62220-1中的标准方法。不透明的钨边测试装置可实现入射X射线的理想阶梯状轮廓。边缘扩展功能是在整个边缘过渡区域的较大区域上进行测量的,从而可以进行精确的MTF测量,包括“低频下降”。该方法已应用于不同的最新X射线成像探测器,计算机射线照相,基于CsI的间接和基于Se的直接平板探测器。将显示MTF测量结果。与不透明边缘设备相反,常用的半透明边缘测试设备会产生散射辐射,使入射X射线轮廓变差,从而导致对MTF的系统性高估。