STUDY OF NEUTRON SCATTERING CONTRIBUTION ON H_p(10) AND H~*(10) CALIBRATION IN THE BRAZILIAN NATIONAL LOW SCATTERING LABORATORY

机译:巴西国家低散射实验室中H_p(10)和H〜*(10)标定的中子散射贡献的研究

获取原文
获取原文并翻译 | 示例
           

摘要

The neutron scattering at the Low Scattering Laboratory of the Brazilian National Neutron Laboratory has been studied using three different methods. The measurements have been done with a traceable standard ~(241)Am-Be from source-to-detector distances of 0.52-3.00 m. The obtained results with the variation distance methods are in agreement. Measurements with a large shadow cone are not worth for larger distances due to overshadowing. As the quantity required in a calibration is the response of the device being calibrated to the scattered neutron component in order to subtract this from the total response, for these purposes, the distance variation method must be used for each device. To quantify absolutely the scattering contribution on the quantity rates of fluence, H_p(10) and H~*(10) in irradiation procedures, a Bonner sphere spectrometer with the shadow cone was employed. The evaluated scattering correction factor value may be employed for a distance of 1.00 m.
机译:巴西国家中子实验室低散射实验室的中子散射已使用三种不同方法进行了研究。使用可追溯的标准〜(241)Am-Be从源到检测器的距离为0.52-3.00 m进行测量。用变化距离法得到的结果是一致的。由于阴影过大,对于较大的距离,不建议使用大阴影锥进行测量。由于校准所需的数量是被校准设备对散射中子成分的响应,以便从总响应中减去它,因此,出于这些目的,必须对每个设备使用距离变化方法。为了绝对量化辐照过程中对通量,H_p(10)和H〜*(10)的数量比率的散射贡献,使用了带阴影锥的Bonner球光谱仪。评估的散射校正因子值可用于1.00 m的距离。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2014年第4期|176-180|共5页
  • 作者单位

    Instituto de Radioprotecao e Dosimetria, Av. Salvador Allende, s, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil;

    Instituto de Radioprotecao e Dosimetria, Av. Salvador Allende, s, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil;

    Instituto de Radioprotecao e Dosimetria, Av. Salvador Allende, s, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil;

    Instituto de Radioprotecao e Dosimetria, Av. Salvador Allende, s, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil;

    Instituto de Radioprotecao e Dosimetria, Av. Salvador Allende, s, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号