A perturbed angular correlation spectrometer for material science studies

机译:用于材料科学研究的扰动角度相关光谱仪

获取原文
获取原文并翻译 | 示例
       

摘要

A four-detector perturbed angular correlation (PAC) spectrometer has been developed with ultra-fast BaF_2 detectors to acquire four coincidence spectra simultaneously, two at 180° and two at 90°. This spectrometer has double efficiency compared to that of a three-detector set-up. Higher efficiency is desirable for PAC studies in solid state physics where large number of coincidences are required to obtain the PAC spectra with good statistics and is particularly useful when the half-lives of the parent probe nuclei used for PAC measurements are ~2-3 days or less as in ~(111)In (2.8 d), ~(99)Mo (2.7 d) and ~(140)La (1.7 d). The performance of the spectrometer has been tested for the HfC>2 monoclinic crystal in the temperature range from 77 to 873 K and for the HfF_4·3H_2O crystal at room temperature. The polycrystalline HfO_2 has been synthesized from Hf metal by heating in air. The hydrated hafnium fluoride has been crystallized by dissolving Hf metal in 40% HF and drying slowly at room temperature.
机译:已经开发了具有超快BaF_2检测器的四探测器扰动角相关(PAC)光谱仪,以同时获取四个重合光谱,两个在180°,两个在90°。与三探测器装置相比,该光谱仪具有两倍的效率。固态物理学中的PAC研究需要更高的效率,因为需要大量重合才能获得具有良好统计数据的PAC光谱,当用于PAC测量的母探针核的半衰期约为2-3天时,该效率特别有用或小于〜(111)In(2.8 d),〜(99)Mo(2.7 d)和〜(140)La(1.7 d)。在室温下,对于HfC> 2单斜晶体在77至873 K的温度范围内以及HfF_4·3H_2O晶体进行了光谱仪的性能测试。通过在空气中加热由Hf金属合成了多晶HfO_2。通过将H金属溶解在40%的HF中并在室温下缓慢干燥,可以使水合氟化ha结晶。