首页> 外文期刊>Nuclear instruments and methods in physics research >Chlorine measurements at the 5MV French AMS national facility ASTER: Associated external uncertainties and comparability with the 6MV DREAMS facility
【24h】

Chlorine measurements at the 5MV French AMS national facility ASTER: Associated external uncertainties and comparability with the 6MV DREAMS facility

机译:5MV法国AMS国家工厂ASTER的氯测量:与6MV DREAMS设备相关的外部不确定性和可比性

获取原文
获取原文并翻译 | 示例
           

摘要

After 6 years of36Cl routine operation, more than 6000 unknown samples have been measured at the 5MV French accelerator mass spectrometry (AMS) national facility ASTER (CEREGE, Aix en Provence). This paper presents the long term behavior of ASTER through the analysis of the measurements of the most used chlorine standards and reference materials, KNSTD1600, SM-Cl-12 and SM-CL-13 over a 46 months’ time period.Comparison of measured chlorine concentrations (both35Cl and36Cl) from ice samples on two AMS facilities operating at 5MV (ASTER) and 6MV (DREAMS, Helmholtz-Zentrum Dresden-Rossendorf) and normalizing to two different reference materials agree within uncertainties making both reference materials (SM-Cl-12 and KNSTD1600) suitable for36Cl measurement at ASTER.
机译:在36Cl常规运行6年后,已在5MV法国加速器质谱(AMS)国家机构ASTER(CEREGE,普罗旺斯地区艾克斯)测量了6000多个未知样品。本文通过分析最常用的氯标准液和标准物质KNSTD1600,SM-Cl-12和SM-CL-13在46个月的时间范围内的测量结果,来介绍ASTER的长期行为。在两个分别运行于5MV(ASTER)和6MV(DREAMS,Helmholtz-Zentrum Dresden-Rossendorf)的AMS设施上从冰样中获得的冰浓度(35Cl和36Cl)并归一化为两种不同的参考物质在不确定性内是一致的,这两种参考物质(SM-Cl-12和KNSTD1600)适用于ASTER的36 Cl测量。