Evaluation of Multiple-Sampling Function used with a Microtek flatbed scanner for Radiation Dosimetry Calibration of EBT2 Film

机译:与Microtek平板扫描仪配合使用的多重采样功能评估EBT2膜的辐射剂量法校准

获取原文
获取原文并翻译 | 示例
           

摘要

The radiochromic EBT2 film is a widely used quality assurance device for radiation therapy. This study evaluated the film calibration performance of the multiple-sampling function, a function of the Scan-Wizard Pro scanning software provided by the manufacturer, when used with Microtek 9800XL plus (9800XL~+) flatbed scanner. By using the PDD method, each one of the eight EBT2 films, four delivered by 290 monitor unit (MU) and four by 88 MU via 6-MV photon beams, was tightly sandwiched in a 30~3-cm~3 water equivalent polystyrene phantom prior to irradiation. Before and after irradiation, all films were scanned using the Microtek 9800XL~+ scanner with five different modes of the multiple-sampling function, which could generate the image with the averaged result of multiple-sampling. The net optical densities (netOD) on the beam central axis of film were assigned to corresponding depth doses for calibration. For each sampling mode with either delivered MU, the depth-dose uncertainty of a single film from repeated scans and that of a single scan of the four films were analyzed. Finally, the calibration error and the combined calibration uncertainty between film determined depth-doses and delivered depth-doses were calculated and evaluated for each sampling mode. All standard deviations and the calibration error were demonstrated to be unrelated to the number of sampling lines. The calibration error of the 2-line and 16-line mode was within 3 cGy and better than that of the other modes. The combined uncertainty of the 2-line mode was the lowest which was generally less than 6 cGy except for the delivered dose around 100 cGy. The evaluation described herein revealed that the EBT2 film calibrated with the 2-line mode has relatively lower error, scanning time and combined uncertianty. Therefore, it is recommended for routine EBT2 film calibration and verification of treatment plans.
机译:辐射变色EBT2膜是用于放射治疗的广泛使用的质量保证装置。这项研究评估了与Microtek 9800XL plus(9800XL〜+)平板扫描仪一起使用时,多重采样功能(制造商提供的Scan-Wizard Pro扫描软件的功能)的胶片校准性能。通过使用PDD方法,将八片EBT2膜中的每一层,其中四层由290个监控单元(MU)输送,四层由88 MU经由6-MV光子束,紧紧地夹在30〜3-cm〜3水当量聚苯乙烯中辐射前幻影。辐照前后,使用具有五种不同模式的多重采样功能的Microtek 9800XL〜+扫描仪对所有胶片进行扫描,可以生成具有多次采样平均结果的图像。将膜的光束中心轴上的净光密度(netOD)分配给相应的深度剂量进行校准。对于具有任一已交付MU的每种采样模式,分析了重复扫描的单张胶片的深度剂量不确定性和四张胶片的单张扫描的深度剂量不确定性。最后,针对每种采样模式,计算并评估了胶片确定的深度剂量和所输送的深度剂量之间的校准误差和组合的校准不确定性。证明所有标准偏差和校准误差与采样线的数量无关。 2线和16线模式的校准误差在3 cGy以内,并且优于其他模式。 2线模式的组合不确定度最低,通常小于6 cGy,但所输送的剂量约为100 cGy。本文所述的评估表明,用2线模式校准的EBT2胶片具有相对较低的误差,扫描时间和合并的不真实性。因此,建议对EBT2胶片进行常规校准和验证治疗计划。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号