摘要

To test the ability of frequency doubling technology (FDT) perimetry to reveal defects in the same field quadrants as detected by standard automated perimetry (SAP). Ninety-two eyes with open-angle glaucoma and documented visual field defects by threshold SAP (Octopus Dynamic strategy) also underwent threshold FDT testing after successfully passing the FDT screening test. All eyes revealed varying stages of SAP defects while only 80 revealed FDT damage: 31:21 eyes in the early field loss stage, 36:35 in the moderate field loss stage, and 25:24 in the severe field loss stage in SAP versus FDT, respectively. SAP was able to detect abnormalities in 74 and 79% of the superotemporal, and inferotemporal quadrants, respectively, while the corresponding FDT figures were 70 and 69% for the same quadrants (P < 0.05 each). With regards to the nasal hemifield, SAP detected defects in 73 and 81% of the superonasal and inferonasal quadrants, respectively, compared to 69 and 66% for FDT (P < 0.001 each). The test duration per individual eye was significantly shorter with FDT than with SAP (P < 0.05). As well as the already established lower sensitivity of FDT compared to SAP, this study also demonstrated the significantly poorer ability of FDT in detecting the same field quadrant defects, especially in the early stages of glaucomatous damage.
机译:为了测试倍频技术(FDT)视野检查法揭示由标准自动视野检查法(SAP)检测到的同一象限中的缺陷的能力。成功通过FDT筛查测试后,有92例具有开角型青光眼且通过阈值SAP(章鱼动态策略)记录的视野缺损也接受了FDT阈值测试。在SAP与FDT相比,所有眼睛都显示出SAP缺陷的不同阶段,而只有80处显示了FDT损坏:早期场损失阶段为31:21眼,中场损失阶段为36:35,重度场丢失阶段为25:24。 , 分别。 SAP能够分别在颞上和下颞象限中检出74%和79%的异常,而同一象限中相应的FDT数字分别为70%和69%(每个P <0.05)。关于鼻半球,SAP在上鼻和下鼻象限分别检测到73%和81%的缺陷,而FDT分别检测到69%和66%的缺陷(每个P <0.001)。 FDT的每只眼睛的测试持续时间明显短于SAP(P <0.05)。除了已经确定的FDT相对于SAP较低的敏感性外,这项研究还证明了FDT检测相同场象限缺陷的能力明显较差,尤其是在青光眼损害的早期阶段。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号