In this paper, accuracy or error in permeability measurement using a combined microstrip line-coaxial conductor method was investigated. The measurement circuit used in this study is composed of a microstrip line (MSL) circuit and a grounded metal pipe with a center conductor passing through it coaxially. A sample is placed between the metal pipe and the center conductor. We evaluated the measurement accuracy for this measurement arrangement with electromagnetic simulation for the case where there are gaps between the sample and the holder. As a result, it has been shown that the normalized errors for this method have similar gap size dependence to the conventional coaxial method, but are about 10 to 20 percent greater than the coaxial method. With a view to improving the measurement accuracy, a correction method for the error is also discussed.
机译:在本文中,研究了使用微带线-同轴电缆组合导体法测量渗透率的准确性或误差。本研究中使用的测量电路由微带线(MSL)电路和接地的金属管组成,中心导体同轴地穿过该金属管。将样品放在金属管和中心导体之间。对于样品和支架之间存在间隙的情况,我们通过电磁仿真评估了此测量装置的测量精度。结果,已经表明,该方法的归一化误差具有与常规同轴方法相似的间隙尺寸依赖性,但是比同轴方法大约10%至20%。为了提高测量精度,还讨论了误差的校正方法。