We evaluate the Scanwell Time Alignment Probe for performing the timing calibration for the LBNL Prostate-Specific PET Camera. We calibrate the time delay correction factors for each detector module in the camera using two methods-using the Time Alignment Probe (which measures the time difference between the probe and each detector module) and using the conventional method (which measures the timing difference between all module-module combinations in the camera). These correction factors, which are quantized in 2 ns steps, are compared on a module-by-module basis. The values are in excellent agreement-of the 80 correction factors, 62 agree exactly, 17 differ by 1 step, and 1 differs by 2 steps. We also measure on-time and off-time counting rates when the two sets of calibration factors are loaded into the camera and find that they agree within statistical error. We conclude that the performance using the Time Alignment Probe and conventional methods are equivalent
机译:我们评估Scanwell时间对准探头,以执行LBNL前列腺专用PET摄像机的定时校准。我们使用两种方法校准相机中每个检测器模块的时间延迟校正因子:使用时间对准探针(用于测量探针与每个检测器模块之间的时间差)和使用常规方法(用于测量所有检测器之间的时间差)相机中的模块-模块组合)。这些校正因子以2 ns的步长进行量化,将逐模块进行比较。这些值非常吻合-80个校正因子,62个精确吻合,17个相差1步,1个相差2步。当将两组校准因子加载到摄像机中并发现它们在统计误差范围内一致时,我们还测量准时和非准时计数率。我们得出的结论是,使用时间对准探针和常规方法的性能相当