摘要

A silicon photomultiplier (SiPM)-based time-of-flight capable PET detector has been integrated with a 70 cm wide-bore 3T MR scanner for simultaneous whole-body imaging (MR750w, GE Healthcare, Waukesha, WI). After insertion of the PET detector, the final PET/MR bore is 60 cm wide (SIGNA PET/MR, GE Healthcare, Waukesha, WI). The MR performance was compared before and after the PET ring insertion. B0 homogeneity, B+1 uniformity of the body coil along with peak B+1, coherent noise, and FBIRN (Function Biomedical Informatics Research Network) tests are used to compare the MR performance. It is shown that B0 homogeneity and coherent noise have not changed according to the system specifications. Peak B+1 is increased by 33% and B+1 inhomogeneity is increased by 4% after PET ring insertion due to a smaller diameter body coil design. The FBIRN test shows similar temporal stability before and after PET ring insertion. Due to a smaller body coil on the PET/MR system, the signal fluctuation to noise ratio (SFNR) and SNR for body receive coil, are improved by ∼40% and ∼160% for Echo Planar Imaging (EPI) and spiral sequences respectively. Comparison using RF- and gradient-intensive clinical sequences shows inserting the PET detectors into the wide-bore MRI has not compromised the MR image quality according to these tests.
机译:基于硅光电倍增管(SiPM)的飞行时间PET检测器已与70厘米宽口径3T MR扫描仪集成在一起,可同时进行全身成像(MR750w,GE Healthcare,Waukesha,WI)。插入PET检测器后,最终的PET / MR孔宽60厘米(SIGNA PET / MR,GE Healthcare,Waukesha,WI)。在插入PET环之前和之后比较MR性能。 B0均匀性,人体线圈的B + 1均匀性以及峰值B + 1,相干噪声和FBIRN(功能性生物医学信息学研究网络)测试用于比较MR性能。结果表明,B0均匀性和相干噪声没有根据系统规格而改变。在插入PET环后,由于较小直径的主体线圈设计,峰值B + 1增加了33%,B + 1不均匀性增加了4%。 FBIRN测试显示,在插入PET环之前和之后,它们具有类似的时间稳定性。由于PET / MR系统上的人体线圈较小,人体回波的信号波动信噪比(SFNR)和SNR对于平面回波成像(EPI)和螺旋序列分别提高了约40%和〜160% 。使用RF和梯度密集型临床序列进行的比较表明,根据这些测试,将PET检测器插入宽口径MRI并不会影响MR图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号