A nasogastric tube is commonly used to pour nutrients into the stomachs of patients who have difficulty swallowing in order to provide them with the required nutrition. However, it is possible to inadvertently insert the tube into the lungs. Thus, a method of obtaining the tip position of the tube is desired. In this paper, we use predata of magnetoimpedance (MI) sensors to estimate the position of the magnet attached to the tube tip using a minimum variance beamformer. Experimental results show that the position estimation error is <10 mm even when the distance from the MI sensors is 150 mm using the predata, whereas the error exceeds 10 mm when the predata is not used. These results suggest that the proposed method is a promising method for hospital applications.
机译:鼻胃管通常用于将营养物质倒入吞咽困难的患者的胃中,以向其提供所需的营养。但是,可能会无意间将管子插入肺部。因此,需要一种获得管的尖端位置的方法。在本文中,我们使用磁阻(MI)传感器的前置数据,使用最小方差波束形成器估算附着在管头上的磁体的位置。实验结果表明,即使使用预数据距MI传感器的距离为150 mm,位置估计误差也小于10 mm,而当不使用预数据时,位置估计误差超过10 mm。这些结果表明,所提出的方法是一种有前途的医院应用方法。