An experimental apparatus was built to measure the time constant of fine wet-bulb thermocouples in air. Rapid response solenoid valves (15 msec response time) were used to control airflow through tubing into which wet-bulb thermocouples were placed. Wet-bulb thermocouples (type T, 0.005 cm diameter) with the tip (bead) covered with a moistened wick were tested. Experiments were performed for air velocities ranging from 1.50 to 2.5 m sec~(-1) (Reynolds number of 2,500 to 4,500) and wet-bulb temperature ranging from 12.5 to 17.4℃. Experimental conditions were selected to simulate human respiratory conditions. Correlation between airflow velocity and wet-bulb thermocouple time constants was not significant at all levels.
机译:建立了一个实验设备来测量空气中细湿球热电偶的时间常数。快速响应电磁阀(响应时间为15毫秒)用于控制通过管道的气流,在管道中放置了湿球热电偶。测试了湿球热电偶(T型,直径0.005厘米),其尖端(焊珠)被湿的芯子覆盖。实验的风速范围为1.50至2.5 m sec〜(-1)(雷诺数为2500至4,500),湿球温度为12.5至17.4℃。选择实验条件以模拟人的呼吸条件。在所有水平上,气流速度和湿球热电偶时间常数之间的相关性均不显着。