Three-dimensional analysis of cervical spine motion: reliability of a computer assisted magnetic tracking device compared to inclinometer

机译:颈椎运动的三维分析:与倾斜仪相比,计算机辅助磁跟踪设备的可靠性

获取原文
获取原文并翻译 | 示例
           

摘要

We aimed to investigate the reliability and reproducibility of a magnetic tracking technique for the assessment of overall cervical spine motion (principal and coupled movements). Ten asymptomatic male volunteers with a mean age of 29.3 years (range 20–37 years) were included in the study. Flexion, extension, left and right lateral bending and left and right axial rotation were measured using a magnetic tracking device (MTD) mounted onto a custom head-piece. For rotational movements in the frontal and sagittal planes the results were compared with the measurements of two standard inclinometers. Intra-observer, inter-observer and intra-instrument reliability was assessed with the intraclass correlation coefficient method. There were no significant differences for all motion measurements between the MTD and the inclinometer. High inter-observer reliability was found in flexion, extension, axial rotation and lateral bending indicating that the testing routine is applicable for different examiners. The intra-observer variability was high in flexion and extension, whereas in lateral bending the reliability coefficients were lower and displayed a fair to good reliability for most of the measurements with the MTD. The results of the MTD were found to be highly comparable with the inclinometer results with an inter-instrument correlation coefficient ranging from 0.88 to 0.99. The MTD is a reliable, reproducible method for three-dimensional motion analysis of the cervical spine and therefore a valuable method both for the clinical assessment of various degenerative and traumatic disorders and as a supplement of different therapeutic procedures and rehabilitation.
机译:我们旨在调查磁性跟踪技术的可靠性和可重复性,以评估整体颈椎运动(主要运动和耦合运动)。这项研究包括十名无症状男性志愿者,平均年龄为29.3岁(20-37岁)。使用安装在定制头戴装置上的磁跟踪设备(MTD)测量屈曲,伸展,左右横向弯曲以及左右轴向旋转。对于额面和矢状面的旋转运动,将结果与两个标准倾角仪的测量结果进行比较。使用组内相关系数法评估观察者内部,观察者之间和仪器内部的可靠性。 MTD和测斜仪之间的所有运动测量值均无显着差异。在屈曲,伸展,轴向旋转和横向弯曲中观察者之间具有很高的可靠性,表明该测试程序适用于不同的检查者。观察者内的挠曲性和伸展性较高,而在横向弯曲中,可靠性系数较低,并且对于使用MTD进行的大多数测量而言,其可靠性都相当好。发现MTD的结果与测斜仪的结果高度可比,仪器之间的相关系数范围为0.88至0.99。 MTD是对颈椎进行三维运动分析的可靠,可重现的方法,因此,对于临床评估各种退行性和创伤性疾病以及作为不同治疗程序和康复的补充,都是一种有价值的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号