In order to develop criteria for the range of data points used for regression line fitting in sweep visually evoked potential (sVEP), which would be objective, clearly specified and give good repeatability and validity, and in order to investigate the effect of luminance on sVEP measurement, visual acuity (VA) and contrast sensitivity (CS) were measured with sVEP in adults aged 17–30 years and children aged 6–8 years. Six to ten participants took part in each experiment. Five criteria (C0–C4) for fitting the regression line were implemented. Test–retest repeatability and validity against psychophysical thresholds at three luminance levels were considered for thresholds and the number of acceptable readings. There were significant effects of criteria (repeated measures ANOVAs, P 0.05). There was little effect of luminance over the 25–100 cd/m2 range used. Overall, C2 performed the best and would be the criterion of choice, giving better repeatability, better validity and more viable plots. Keywords sVEP - Visually evoked potential - Contrast sensitivity - Visual acuity
机译:为了制定用于在扫视视觉诱发电位(sVEP)中进行回归线拟合的数据点范围的标准,这将是客观的,明确指定的并且具有良好的可重复性和有效性,并且为了研究亮度对sVEP的影响使用sVEP对17至30岁的成年人和6至8岁的儿童进行测量,视敏度(VA)和对比敏感度(CS)。每个实验有六到十名参与者。实施了五项用于拟合回归线的标准(C0–C4)。考虑阈值和可接受读数的数量,针对三种亮度水平下针对心理阈值的重测重复性和有效性。标准有显着影响(重复测量方差分析,P 0.05)。在使用的25-100 cd / m 2 范围内,亮度几乎没有影响。总体而言,C2表现最好,将成为选择的标准,从而提供更好的可重复性,更好的有效性和更可行的绘图。 sVEP-视觉诱发电位-对比敏感度-视敏度

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号