首页> 外文期刊>International Journal of Medical Physics, Clinical Engineering and Radiation Oncology >Comparative Studies of Absolute Dose in Water Phantom, Solid Water Phantom and MatriXX with MULTICube Phantom
【24h】

Comparative Studies of Absolute Dose in Water Phantom, Solid Water Phantom and MatriXX with MULTICube Phantom

机译:多伏模型水体模,固体水体模和基质绝对剂量的比较研究

获取原文
           

摘要

Introduction: Radiotherapy is one of the important treatment modalities in cancer treatment. To maintain the treatment procedure accuracy, the phantom is an essential tool for absolute dosimetry conformation and Quality Assurance routine cheek up. This work aims to study the absorbed dose of various phantoms and hence to make a comparison of the result with the IAEA recommendation (TRS 398) for daily QA of Linac. Materials and Methods: The experiment has been done at the Institute of Nuclear Medical Physics (INMP), AERE, Savar, Dhaka, under the Bangladesh Atomic Energy Commission. For external beam radiation, 6 MV and 15 MV photon beams of Varian Clinac iX Linear Accelerator (Linac) were used. One dimensional (1D) water phantom, solid water phantom, and MatriXX with MULTICube phantom and associate accessories were used to experiment. Results: We have measured and compared the absorbed dose data of the phantoms. The variation of sold water phantom from the 1D water phantom is +2.8% at 6 MV and +3.5% at 15 MV. The variation MatriXX with MULTIcube phantom from the 1D water phantom is +8.0% at 6 MV and +3.2% at 15 MV. This study revealed that the 1D water phantom was the best absolute dose conformation among the other phantoms and the deviation was within the acceptable limit (±5%), except MatriXX with MULTICube Phantom for low energy beam (8%). Conclusion: It was observed that the accuracy of dose estimation was better in the 1D water phantom rather than the other two. It is also known that the 1D water phantom is low cost but needs a long time to set up for the experimental arrangement. Solid water or MatriXX with MULTICube phantom can be used to overcome this problem, which takes only a few minutes for setup and is comparatively faster than 1D water phantom.
机译:介绍:放射疗法是癌症治疗中的重要治疗方式之一。为了保持治疗程序准确性,幻影是绝对剂量化构象和质量保证常规脸颊的重要工具。这项工作旨在研究各种幽灵的吸收剂量,从而比较Linac日常QA的IAEA建议(TRS 398)的结果。 材料和方法:在孟加拉国原子能委员会下,核医学物理研究所(inmp),艾默,萨马尔,达卡,达卡队进行了实验。对于外部光束辐射,使用6mV和15个MV光子梁的Varian Clinac IX线性加速器(LINAC)。使用多尺幻影和助理配件的一维(1D)水幻影,固体水幻影和矩阵X实验。 结果:我们已经测量并比较了幻影的吸收剂量数据。从1D水幻影的出售水体模的变化在6mV下+ 2.8%,+ 3.5%在15 mV下。具有来自1D水体幻影的多伏幻影的变化基质为+ 8.0%,在6mV下为+ 3.2%,15 mV。该研究表明,1D水体幻影是其他模糊的最佳的绝对剂量构象,并且偏差在可接受的极限(±5%)内,除了用于低能量束的多伏虚线(8%)的Matrixx。 结论:观察到剂量估计的准确性在1D水幻影中而不是另外两个更好。还众所周知,1D水幻影是低成本,但需要长时间设置实验布置。具有多尺子幻像的固体水或矩阵可用于克服该问题,这仅需要几分钟的设置,并且比1D水幻影相对较快。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号