首页> 外文期刊>Practical Laboratory Medicine >Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
【24h】

Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers

机译:新生儿血红蛋白对Gem Premier 4000血气分析仪进行全血胆红素测量的准确性

获取原文
       

摘要

ObjectivesWhole blood bilirubin measured on blood gas analyzers is accepted by physicians in neonatal hyperbilirubinemia management since it requires a small sample volume. The accuracy of bilirubin measurement on blood gas analyzers is instrument dependent and remains controversial.Design and MethodsBilirubin in adult and umbilical cord whole blood samples, spiked with an unconjugated bilirubin standard, and non-spiked adult plasma samples was measured on a blood gas analyzer (GEM 4000) and a Core Laboratory Chemistry analyzer (Architect c16000) respectively. We also investigated the linear regression for neonatal and adult hemoglobin measured on the blood gas analyzer and the Core Laboratory hematology analyzer (Alinity h-Series).ResultsPlasma bilirubin measured on the blood gas analyzer and the chemistry analyzer was statistically identical. Adult whole blood bilirubin showed slightly increased proportional bias. When umbilical cord whole blood samples were used, the Deming regression showed GEM bilirubin =1.233(Architect) (95% CI 1.199 ~ 1.266)-44.43??μmol/L (95% CI -53.6??~???35.2). The regression was significantly different from that in plasma (p??50??μmol/L showed that the bias% was above laboratory standards. In addition, the regression of neonatal hemoglobin measurement between the GEM and the Alinity was significantly different from adult hemoglobin (p??
机译:在血气分析仪上测量的目标血胆红素被新生儿高胆管素血症管理中的医生接受,因为它需要小的样品体积。胆红素测量对血气分析仪的准确性是仪器依赖性,并且仍然存在争议。成人和脐带的嗜合素霉素在成人和脐带的全血样品中,用无缀合的胆红素标准标准,并在血气分析仪上测量非尖刺的成年等离子体样品(宝石4000)和核心实验室化学分析仪(Architect C16000)。我们还研究了在血气分析仪上测量的新生儿和成人血红蛋白的线性回归和核心实验室血液分析仪(ALINITITE H系列)。在血气分析仪上测量的胆红素,化学分析仪在统计上相同。成人全血胆红素显示出比例偏差略微增加。当使用脐带全血样品时,脱垂回归显示宝石胆红素= 1.233(建筑师)(95%CI 1.199〜1.266)-44.43 ??μmol/ L(95%CI -53.6 ??〜??? 35.2)。回归显着与血浆(p ?? <0.001)或成人全血(p≤0.001)样品显着不同。 36.1%具有胆红素水平>50Ωμmol/ L的新生儿样品表明,偏差%高于实验室标准。此外,宝石和异性之间的新生儿血红蛋白测量的回归与成年血红蛋白有显着不同(P ??

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号