An improved method for measurement of testosterone in human plasma and saliva by ultra-performance liquid chromatography-tandem mass spectrometry

机译:通过超高效液相色谱 - 串联质谱法测定人血浆和唾液中睾酮和唾液的改进方法

获取原文
       

摘要

The aim of the study was to develop and validate a practical assay of clinically relevant testosterone levels in human plasma and saliva. We performed ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) analysis on Atlantis dC18 steel column using a mobile phase of 2-mM ammonium acetate and acetonitrile (20:80, v: v) that was delivered at 0.3 ml/min. After adding d3-testosterone as an internal standard (IS), we extracted plasma and salivary samples with methyl tert-butyl ether. Mass spectrometry was performed in electrospray positive-ion mode. Targeted ion transitions were examined at m/z 289.18 → 97.04 and 292.24 → 97.04 for testosterone and IS, respectively. We validated the method according to the US Food and Drug Administration guidelines. Elution times for testosterone and IS were both around 1.35 min. Testosterone level was linearly associated (rsup2/sup = 0.9975 and 0.9958) with peak area ratio of testosterone to IS between 0.5–50 ng/ml and 10–400 pg/ml in plasma and saliva, respectively. The coefficient of variation and bias were ≤12.6% and ≤±12.1% in plasma and ≤10.2% and ≤±5.3% in saliva. The extraction recovery of testosterone was ≥92% from plasma and ≥94% from saliva. Testosterone was stable (≥91%) for 24 h at room temperature and for 8 weeks at ?20°C in both plasma and salivary samples. We report a simple, validated, UPLC-MS/MS assay that can be used to determine clinically relevant levels of testosterone in human plasma and saliva.
机译:该研究的目的是发展和验证人类血浆和唾液中临床相关睾酮水平的实际测定。我们使用2mM乙酸铵和乙腈(20:80,V:V)的流动相,对亚特兰蒂斯DC18钢柱进行超高效液相色谱 - 串联质谱(UPLC-MS / MS)分析。在0.3时递送ml / min。添加D3睾酮作为内标(是)后,我们用甲基叔丁基醚提取血浆和唾液样品。在电喷雾正离子模式下进行质谱。在M / Z 289.18→97.04和292.24→97.04中检查靶离子转变,分别为睾酮。我们验证了根据美国食品和药物管理指南的方法。睾酮的洗脱时间均约为1.35分钟。睾酮水平线性相关(R 2 = 0.9975和0.9958),睾酮峰面积比在血浆和唾液中的0.5-50ng / ml和10-400pg / ml之间。变异系数和偏差系数≤12.6%,血浆中≤12.1%,唾液中≤10.2%,≤±5.3%。睾酮的提取恢复≥92%从血浆和唾液中≥94%。睾酮在室温下稳定(≥91%)24小时,在血浆和唾液样品中在20℃下持续8周。我们报告了一种简单,验证的UPLC-MS / MS测定,可用于确定人血浆和唾液中的睾丸激素的临床相关水平。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号