首页> 外文期刊>Nigerian Medical Journal >Comparative assessment of the goldmann applanation and noncontact tonometers in intraocular pressure measurements in a sample of glaucoma patients in the Cape Coast Metropolis, Ghana
【24h】

Comparative assessment of the goldmann applanation and noncontact tonometers in intraocular pressure measurements in a sample of glaucoma patients in the Cape Coast Metropolis, Ghana

机译:Goldmann施加和非接触性质计在普纳海岸大都市青光眼患者样本中的高曼施加和非接触眼压计的比较评估

获取原文
           

摘要

Objectives: The objective of the study was to explore the usage of the Goldmann applanation tonometry and noncontact tonometry interchangeably in the measurement of intraocular pressure (IOP) in glaucoma patients. Materials and Methods: The study involved 441 clinically diagnosed glaucoma patients receiving care at a referral facility. IOP measurements were obtained using both the Noncontact tonometer and Goldmann applanation tonometer The repeatability of the measures was analyzed by comparing the repeated measures of the devices using paired t-test and calculating the correlation coefficient. A Bland–Altman analysis was used to determine the limits of agreement between the two procedures. Results: There were 271 (61.5%) males and 170 (38.5%) females and their age ranged from 18 to 73 years (mean age = 49.37; standard deviation ± 14.81 years). The findings of the study showed significantly lower readings (P 0.001) of the GAT (right eye = 17.40 ± 7.48 mmHg; left eye = 16.80 ± 7.49 mmHg) compared to the NCT (right eye = 20.15 ± 8.30 mmHg; left eye = 19.74 ± 8.31 mmHg). There was a strong positive correlation between the GAT and NCT findings in the right eye (r = 0.871, n = 441, P 0.001) and in the left eye (r = 0.887, n = 441, P 0.001). There was a wide limit of agreement between NCT and GAT measurements. Conclusion: There was statistically significant higher measures obtained with NCT than the GAT but did not exceed the allowable inter-device difference. There was a strong positive correlation between GAT and NCT measurements. However, it is strongly recommended that these devices are not used interchangeably in the monitoring of IOP in glaucoma due to the wide range of limits of agreement.
机译:目的:该研究的目的是探讨Goldmann Applanation Tonoxatrationry和Noncontact Tononetry的使用可互动在青光眼患者的眼压(IOP)测量中。材料与方法:该研究涉及441临床诊断的肺炎患者在推荐设施接受护理。使用非接触眼压计和Goldmann施换眼压计获得IOP测量,通过比较使用配对T检验并计算相关系数的装置的重复测量来分析措施的可重复性。使用Bland-Altman分析来确定两种程序之间的协议限制。结果:271例(61.5%)男性和170名(38.5%)女性,其年龄从18至73岁(平均年龄= 49.37;标准差±14.81岁)。该研究的结果显示出读数(右眼= 17.40±7.48mmHg)显着降低读数(P <0.001);与NCT相比,左眼= 16.80±7.49 mmHg)(右眼= 20.15±8.30mmhg;左眼= 19.74±8.31 mmhg)。右眼(r = 0.871,n = 441,p <0.001)和左眼(r = 0.887,n = 441,p <0.001),GAT和NCT结果之间存在强烈的正相关性。 NCT和GAT测量之间存在很大的协议。结论:统计上显着的较高措施,比GAT比GAT获得,但没有超过允许的局部间差异。 GAT和NCT测量之间存在强烈的正相关。然而,强烈建议,由于各种协议限制,这些器件在葡萄糖中的监测中不可互换地使用。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号