Comparative Evaluation of Accuracy of Pulse Oximeters and Factors Affecting Their Performance in a Tertiary Intensive Care Unit OC05-OC08

机译:在三级重症监护病房OC05-OC08中比较脉搏血氧仪准确性和影响其性能的因素的比较评估

获取原文
           

摘要

Introduction: Pulse oximetry is a widely used tool, unfortunately there is a paucity of data investigating its accuracy in Intensive Care Units (ICU) and if they are able to meet mandated FDA criteria as claimed by them in critically ill patients.Aim: To assess bias, precision and accuracy of pulse oximeters used in ICU and factors affecting them.Materials and Methods: A prospective cohort study, including 129 patients admitted to the ICU of a tertiary referral centre. Pulse oximetry and blood gas were done simultaneously. Pulse oximetry was done using two pulse oximetres: Nonin and Philips. All physiological variables like haemoglobin, lactate, use of vasopressors and blood pressure were recorded. Bland Altman curves were constructed to determine bias and limits of agreement. Effect of physiological variables on bias and difference between performance characteristics of bias was determined using SPSS.Results: Pulse oximetry overestimated arterial oxygen saturation (SaO2) by 1.44%. There was negative correlation between bias and SaO2 (r=-0.32) and positive correlation with lactate (r=0.16). The Philips pulse oximeter had significant higher bias and variability than Nonin pulse oximeter. (2.49±2.99 versus 0.46±1.68, mean difference = 1.98, 95% C.I. = 1.53 ? 2.43, p-value <0.001).Conclusion: Pulse oximetry overestimates SaO2. Bias tends to increase with rising lactate and hypoxia. There is heterogeneity in performance of various pulse oximetry devices in ICU.
机译:简介:脉搏血氧饱和度测定法是一种广泛使用的工具,不幸的是,缺乏数据来研究其在重症监护病房(ICU)中的准确性以及是否能够满足危重患者声称的FDA强制性标准。材料和方法:一项前瞻性队列研究,包括129名三级转诊中心的ICU患者。脉搏血氧仪和血气同时进行。脉搏血氧饱和度测定使用两种脉搏肟:Nonin和Philips。记录所有生理变量,如血红蛋白,乳酸,使用升压药和血压。构造Bland Altman曲线以确定偏差和一致性极限。使用SPSS确定了生理变量对偏倚的影响以及偏倚的性能特征之间的差异。结果:脉搏血氧饱和度高估了动脉血氧饱和度(SaO2)1.44%。偏倚与SaO2之间呈负相关(r = -0.32),而与乳酸呈正相关(r = 0.16)。飞利浦脉搏血氧仪比Nonin脉搏血氧仪具有更高的偏差和可变性。 (2.49±2.99比0.46±1.68,平均差异= 1.98,95%C.I. = 1.53〜2.43,p值<0.001)结论:脉搏血氧饱和度高估了SaO2。偏倚往往随着乳酸和缺氧的增加而增加。 ICU中各种脉搏血氧仪的性能存在异质性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号