首页> 外文期刊>Indian Journal of Science and Technology >The Application of a Water Filter to Reduce Radiation during Bone Mineral Densitometry (BMD)
【24h】

The Application of a Water Filter to Reduce Radiation during Bone Mineral Densitometry (BMD)

机译:滤水器在骨矿物质密度测定(BMD)过程中减少辐射的应用

获取原文
       

摘要

In order to provide a set of basic data that can help to minimize the radiation during bone mineral densitometry and to acquire an optimal image for diagnosis, a water filter was used with a glass dosimeter to acquire an image, and the T-score and Z-score between the image and the radiation amount were measured.Radiation amount was 18+-0.36 μGy for the lens, 17+-0.42 μGy for the thyroid, 18+-0.33 μGy for the breasts, 19+-0.41 μGy for the heart and 18+-0.35 μGy for the reproductive organs. When the water filter was applied by 1cm, 2cm, 3cm, 4cm and 5cm, the radiation decreased by 11.1%. The acquisition of the same T-score and Z score before and after the application of a water filter was to 1cm and for 2cm or more; the same values for T-score and Z-score could not be acquired.The results are expected to provide a basic set of data in forecasting the application of water filters during BMD and determining the test method, leading to reduced radiation.
机译:为了提供一组基本数据,可以帮助最大程度地减少骨矿物质密度测定过程中的辐射并获得最佳的诊断图像,水过滤器与玻璃剂量计一起使用以获取图像,T分数和Z测量图像与放射线量之间的分数-镜片的放射线量为18 + -0.36μGy,甲状腺的放射线量为17 + -0.42μGy,乳房的放射线量为18 + -0.33μGy,心脏的放射线量为19 + -0.41μGy生殖器官为18 + -0.35μGy。当将滤水器应用1cm,2cm,3cm,4cm和5cm时,辐射降低了11.1%。在使用滤水器之前和之后,获得相同的T分数和Z分数为1cm和2cm或更长;预期结果将为预测BMD期间滤水器的应用和确定测试方法提供一组基础数据,从而减少辐射。