首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Effect of pH and NaCl on measurements of ionized calcium in matrices of serum and human albumin with a new calcium-selective electrode.
【24h】

Effect of pH and NaCl on measurements of ionized calcium in matrices of serum and human albumin with a new calcium-selective electrode.

机译:pH和NaCl对使用新型钙选择电极测量血清和人白蛋白基质中离子钙的影响。

获取原文
           

摘要

A neutral ligand ion-exchanger type of ion-selective electrode is used in the current version of the Orion Model SS-20 calcium analyzer. We evaluated it over the pH range 4.0 to 8.5, using albumin solutions containing 150 mmol/L NaCl and pooled patients' serum. Within this pH range response was nernstian, and the relationship between potential (in millivolts) and log [Ca2+] at each pH was linear. Variation in Ca2+ with increasing ionic strength (because of added NaCl) was not significantly different between standards and albumin solutions; it decreased in both. In contrast, the proportion of calcium in serum increased with increasing NaCl concentrations. Ca2+ was not bound in whole serum albumin at a pH less than about 3.9. We observed a substantial difference between the degree of Ca2+ binding in solutions of fatty-acid-free human albumin and serum.
机译:当前版本的Orion Model SS-20钙分析仪使用中性配体离子交换剂类型的离子选择电极。我们使用包含150 mmol / L NaCl的白蛋白溶液和合并的患者血清在4.0至8.5的pH范围内对其进行了评估。在此pH范围内,响应是神经性的,并且每个pH下的电势(以毫伏为单位)和log [Ca2 +]之间的关系是线性的。随着离子强度的增加(由于添加了NaCl),Ca2 +的变化在标准溶液和白蛋白溶液之间没有显着差异。两者都减少了。相反,血清中钙的比例随NaCl浓度的增加而增加。在小于约3.9的pH值下,Ca2 +未结合在全血清白蛋白中。我们观察到在不含脂肪酸的人白蛋白和血清溶液中Ca2 +结合程度之间存在实质性差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号