首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >A New and Rapid Method for the Determination of Glucose by Measurement of Rate of Oxygen Consumption
【24h】

A New and Rapid Method for the Determination of Glucose by Measurement of Rate of Oxygen Consumption

机译:一种测定耗氧率的快速测定葡萄糖的新方法

获取原文
           

摘要

Glucose levels in serum, plasma, and urine are determined rapidly and conveniently by a new glucose oxidase method employing a polarographic oxygen sensor with a circuit modified to record the rate of oxygen consumption. The maximum apparent rate of oxygen consumption relative to the rate obtained with a glucose standard provides a direct measure of the glucose level in the sample; results are obtainable within 20 sec. after sample (100 μl.) addition and within 3 min. after a blood sample is withdrawn from a patient. Interferences associated with prior colorimetric glucose oxidase methods are avoided by measuring oxygen consumption instead of hydrogen peroxide formation. The method is described and results are presented showing a standard deviation of less than 1.5% on replicate determinations and a bias of 1% with respect to data obtained on the same samples by the automated ferricyanide method.
机译:通过一种新的葡萄糖氧化酶方法可以快速,方便地确定血清,血浆和尿液中的葡萄糖水平,该方法采用了极谱氧传感器,并修改了电路以记录耗氧率。相对于用葡萄糖标准液获得的最大耗氧量,表观耗氧率可以直接测量样品中的葡萄糖水平。结果可在20秒内获得。加入样品(100μl)后3分钟内从患者身上抽取血液样本后。通过测量耗氧量而不是过氧化氢的形成,可以避免与现有比色葡萄糖氧化酶方法相关的干扰。描述了该方法,并给出了结果,表明重复测定的标准偏差小于1.5%,相对于通过自动铁氰化物方法在相同样品上获得的数据而言,偏差为1%。