首页> 外文期刊>Journal of Medical Physics/Association of Medical Physicists of India >Imaging and Dosimetric Study on Direct Flat-Panel Detector-Based Digital Mammography System
【24h】

Imaging and Dosimetric Study on Direct Flat-Panel Detector-Based Digital Mammography System

机译:基于直接平板探测器的数字乳腺X线摄影系统的成像和剂量学研究

获取原文
           

摘要

Introduction: Image quality of digital mammography system is generally defined by three primary physical parameters, namely, contrast, resolution, and noise. Quantification of these metrics can be done by measuring objective image quality parameters defined as contrast-to-noise ratio (CNR), modulation transfer function (MTF), and noise power spectra (NPS). Materials and Methods: In the present study, various imaging metrics such as CNR, contrast detail resolution, MTF, and NPS were evaluated for a direct flat-panel detector-based digital mammography system following the European Guidelines. Furthermore, system performance relating to both image quality and doses were evaluated using figure of merit (FOM) in terms of CNR2/mean glandular dose (MGD) under automatic exposure control (AEC) and clinically used OPDOSE operating mode. Results and Conclusion: Under AEC mode, FOM values for the 4.5 cm thick BARC polymethyl methacrylate (PMMA) phantom were found to be 15.02, 15.88, and 19.82 at Mo/Mo, Mo/Rh, and W/Rh target/filter (T/F), respectively. Under OPDOSE mode, FOM values were found to 65.32, 11.80, and 1.14 for the BARC PMMA phantom thickness of 2, 4.5, and 8 cm, respectively. Under OPDOSE mode, the calculated MGD values for three Computerized Imaging Reference Systems slab phantoms having total thickness of 7 cm were observed to be 3.03, 2.32, and 1.75 mGy with glandular/adipose tissue compositions of 70/30, 50/50, and 30/70, respectively, whereas for the 2–8-cm thick BARC PMMA phantom, the calculated MGDs were found to be in the range of 0.57–3.32 mGy. All the calculated MGDs values were found to be lower than the acceptable level of dose limits provided in European Guidelines.
机译:简介:数字乳房X线摄影系统的图像质量通常由三个主要物理参数定义,即对比度,分辨率和噪声。这些指标的量化可以通过测量客观图像质量参数来完成,这些参数定义为对比噪声比(CNR),调制传递函数(MTF)和噪声功率谱(NPS)。材料和方法:在本研究中,按照欧洲指南,对直接基于平板探测器的数字乳腺摄影系统评估了各种成像指标,例如CNR,对比度细节分辨率,MTF和NPS。此外,在自动曝光控制(AEC)和临床使用的OPDOSE操作模式下,使用品质因数(FOM)根据CNR2 /平均腺体剂量(MGD)评估了与图像质量和剂量相关的系统性能。结果与结论:在AEC模式下,发现在Mo / Mo,Mo / Rh和W / Rh目标/滤光片(T处,4.5 cm厚的BARC聚甲基丙烯酸甲酯(PMMA)幻像的FOM值为15.02、15.88和19.82。 / F)。在OPDOSE模式下,对于BARC PMMA幻影厚度分别为2、4.5和8 cm的FOM值分别为65.32、11.80和1.14。在OPDOSE模式下,观察到三个计算机成像参考系统平板幻影的总MGD值分别为7 cm,分别为3.03、2.32和1.75 mGy,腺/脂肪组织组成为70 / 30、50 / 50和30 / 70,而对于2–8-cm厚的BARC PMMA体模,计算出的MGD被发现在0.57–3.32 mGy的范围内。发现所有计算出的MGDs值均低于欧洲指南中提供的可接受的剂量限值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号